Literature DB >> 28034811

The Use of Intraoperative Electromyogram During Spinal Cord Stimulator Placement Surgery: A Case Series.

Nathan Schoen1, Lee Onn Chieng1, Karthik Madhavan2, Walter J Jermakowicz1, Steven Vanni1.   

Abstract

BACKGROUND: Spinal cord stimulation (SCS) is an efficacious treatment for various chronic pain syndromes culminating predominantly into spinal nerves. To improve intraoperative electrode placement, several groups have advocated the use of intraoperative neuromonitoring for localization of the spinal cord midline. In our study we present the case series of patients undergoing stimulator placement with consistent electromyographic intraoperative testing, with an emphasis on examining reoperation rates and complications.
METHODS: After approval from the institutional review board, we retrospectively collected data on standard demographics, preoperative diagnoses, prior spine surgeries, electrode manufacturer, blood loss, complications, and patient outcome.
RESULTS: The study included 103 patients with an average age of 60.7 years. Of these, 72 (69.9%) had prior spine surgery, which was associated with higher rate of reoperation (P = 0.019). The mean latency between initial SCS implantation and revision surgery ± SD was 14.6 ± 17.2 months, with a median time of 280 days. There was a 13.6% complication rate. Common complications that lead to reoperation included migrated electrode or failed generator.
CONCLUSIONS: Our retrospective chart review of 103 patients indicates that patients receiving SCS implantation in conjunction with electromyographic monitoring have low complication rates and rarely return to the operating room for electrode repositioning or removal.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Chronic pain; Electromyogram; Failed back surgery syndrome; Spinal cord stimulation

Mesh:

Year:  2016        PMID: 28034811     DOI: 10.1016/j.wneu.2016.12.077

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104