Hee Jae Huh1, Jong Eun Park1, Ji Youn Kim2, Sun Ae Yun2, Myoung Keun Lee1, Nam Yong Lee1, Jong Won Kim1, Chang Seok Ki3.
Abstract
There has been increasing interest in standardized and quantitative Epstein-Barr virus (EBV) DNA testing for the management of EBV disease. We evaluated the performance of the Real-Q EBV Quantification Kit (BioSewoom, Korea) in whole blood (WB). Nucleic acid extraction and real-time PCR were performed by using the MagNA Pure 96 (Roche Diagnostics, Germany) and 7500 Fast real-time PCR system (Applied Biosystems, USA), respectively. Assay sensitivity, linearity, and conversion factor were determined by using the World Health Organization international standard diluted in EBV-negative WB. We used 81 WB clinical specimens to compare performance of the Real-Q EBV Quantification Kit and artus EBV RG PCR Kit (Qiagen, Germany). The limit of detection (LOD) and limit of quantification (LOQ) for the Real-Q kit were 453 and 750 IU/mL, respectively. The conversion factor from EBV genomic copies to IU was 0.62. The linear range of the assay was from 750 to 10⁶ IU/mL. Viral load values measured with the Real-Q assay were on average 0.54 log₁₀ copies/mL higher than those measured with the artus assay. The Real-Q assay offered good analytical performance for EBV DNA quantification in WB.Entities:
Keywords: Epstein-Barr virus; Performance; Quantification; Real-time PCR; Whole blood
Mesh:
Substances:
Year: 2017 PMID: 28029001 PMCID: PMC5203992 DOI: 10.3343/alm.2017.37.2.147
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Analytical performance of the Real-Q EBV Quantification Kit
| Analytical performance | Result |
|---|---|
| Limit of detection | 453 IU/mL (2.66 log10 IU/mL) |
| Limit of quantification | 750 IU/mL (2.88 log10 IU/mL) |
| Linear range | 2.88–6.00 log10 IU/mL |
| Conversion factor | 0.62 IU/copy |
Fig. 1Linearity of the Real-Q EBV Quantification Kit.
Fig. 2Comparison of clinical specimens tested with the Real-Q and artus assays. A total of 38 specimens with EBV DNA load above the limit of quantification in both assays were analyzed. (A) A Bland-Altman plot comparing the artus and Real-Q assays. The solid line represents the mean of difference between copy number values obtained by the Real-Q and the artus assays; dashed lines indicate the 95% confidence interval of the mean of difference. (B) The Passing-Bablok regression analysis. The regression line (solid line) and the 95% confidence interval (dashed lines) are displayed.
Abbreviation: EBV, Epstein-Barr virus.