Literature DB >> 28027761

Radiation dose reduction during transjugular intrahepatic portosystemic shunt implantation using a new imaging technology.

C Spink1, M Avanesov2, T Schmidt3, M Grass4, G Schoen5, G Adam2, P Bannas2, A Koops2.   

Abstract

OBJECTIVE: To compare patient radiation dose in patients undergoing transjugular intrahepatic portosystemic shunt (TIPS) implantation before and after an imaging-processing technology upgrade.
METHODS: In our retrospective single-center-study, cumulative air kerma (AK), cumulative dose area product (DAP), total fluoroscopy time and contrast agent were collected from an age- and BMI-matched collective of 108 patients undergoing TIPS implantation. 54 procedures were performed before and 54 after the technology upgrade. Mean values were calculated and compared using two-tailed t-tests. Two blinded, independent readers assessed DSA image quality using a four-rank likert scale and the Wilcoxcon test.
RESULTS: The new technology demonstrated a significant reduction of 57% of mean DAP (402.8 vs. 173.3Gycm2, p<0.001) and a significant reduction of 58% of mean AK (1.7 vs. 0.7Gy, p<0.001) compared to the precursor technology. Time of fluoroscopy (26.4 vs. 27.8min, p=0.45) and amount of contrast agent (109.4 vs. 114.9ml, p=0.62) did not differ significantly between the two groups. The DSA image quality of the new technology was not inferior (2.66 vs. 2.77, p=0.56).
CONCLUSIONS: In our study the new imaging technology halved radiation dose in patients undergoing TIPS maintaining sufficient image quality without a significant increase in radiation time or contrast consumption.
Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Image-processing technology; Interventional radiology; Liver; Radiation dosage; Transjugular intrahepatic portosystemic shunt (TIPS)

Mesh:

Year:  2016        PMID: 28027761     DOI: 10.1016/j.ejrad.2016.11.028

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Tantalum-specific contrast-to-noise ratio or conventional detector dose-driven exposure control in angiography: radiation dose and image quality evaluation in a porcine model.

Authors:  Thomas Werncke; Timo Christian Meine; Jan B Hinrichs; Sabine K Maschke; Lena Sophie Becker; Inga Brüsch; Regina Rumpel; Frank K Wacker; Bernhard C Meyer
Journal:  Eur Radiol Exp       Date:  2022-05-17

2.  Radiation dose reduction during adrenal vein sampling using a new angiographic imaging technology.

Authors:  Clemens Spink; Maxim Avanesov; Alexander Lenz; Frank Oliver Henes; Lennart Well; Thomas Schmidt; Gerhard Adam; Harald Ittrich; Peter Bannas
Journal:  Sci Rep       Date:  2022-04-12       Impact factor: 4.379

  2 in total