Abstract
Chemiluminescence (CL) has been widely applied in many fields, but it is rarely used in a very simple, economical but effective way. In this work, for the first time, the CL cloth-based glucose test sensors (CCGTSs) are developed as a new class of CL glucose sensors, with no need for complicated, expensive device fabrication and peripheral equipment. When integrated with desirable hydrophobic barrier in the flow channel and gravity/capillary flow induced by a difference in height between the loading zone and the detection zone, a single cloth-based device can perform the whole CL process involving two steps of enzyme reactions. The wax screen-printing approach is used to fabricate ultra-cheap CCGTSs, the glucose detection involves the enzymatic oxidation of glucose to gluconic acid and H2O2 followed by the horseradish peroxidase (HRP)-catalyzed oxidation of luminol by H2O2, and the emitted CL signals are heightened with p-iodophenol (PIP) and imaged using an inexpensive, portable CCD camera. Under optimized conditions, glucose can be determined over the range of 0.1-100mM, with a detection limit of 0.0948mM and an analysis time of less than 5.5min. Finally, the applicability and validity of the CCGTSs are demonstrated for the measurement of glucose in clinical urine and serum samples. Thus, the proposed sensors could provide great promise in applications in many areas, and may facilitate the achievement of point-of-care testing.Entities:
Keywords: Chemiluminescence; Glucose detection; Gravity/capillary flow; Microfluidic cloth-based analytical devices; Wax screen-printing
Mesh:
Substances:
Year: 2016 PMID: 28024284 DOI: 10.1016/j.bios.2016.12.004
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618