Microplasma-based flowing atmospheric-pressure afterglow (FAPA) source for ambient desorption-ionization mass spectrometry.

Offer M Zeiri1, Andrew P Storey1, Steven J Ray2, Gary M Hieftje1.   

Abstract

A new direct-current microplasma-based flowing atmospheric pressure afterglow (FAPA) source was developed for use in ambient desorption-ionization mass spectrometry. The annular-shaped microplasma is formed in helium between two concentric stainless-steel capillaries that are separated by an alumina tube. Current-voltage characterization of the source shows that this version of the FAPA operates in the normal glow-discharge regime. A glass surface placed in the path of the helium afterglow reaches temperatures of up to approximately 400 °C; the temperature varies with distance from the source and helium flow rate through the source. Solid, liquid, and vapor samples were examined by means of a time-of-flight mass spectrometer. Results suggest that ionization occurs mainly through protonation, with only a small amount of fragmentation and adduct formation. The mass range of the source was shown to extend up to at least m/z 2722 for singly charged species. Limits of detection for several small organic molecules were in the sub-picomole range. Examination of competitive ionization revealed that signal suppression occurs only at high (mM) concentrations of competing substances.
Copyright © 2016 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ambient desorption-ionization; Ion source; Mass spectrometry; Microplasma

Year:  2016        PMID: 28010838     DOI: 10.1016/j.aca.2016.10.042

Source DB:  PubMed          Journal:  Anal Chim Acta        ISSN: 0003-2670            Impact factor:   6.558


  1 in total

1.  Glow flow ionization mass spectrometry of small molecules. A comparison of a glow flow ionization source ('GlowFlow') with electrospray ionization and atmospheric pressure chemical ionization.

Authors:  Rhodri N Owen; Stevan Bajic; Steven L Kelly; Michael R Morris; A Gareth Brenton
Journal:  Rapid Commun Mass Spectrom       Date:  2022-08-15       Impact factor: 2.586

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.