| Literature DB >> 28009337 |
Z H Ibrahim1, G Chari2, S Abdel Baki3, V Bronshtein1, M R Kim4, J Weedon5, J Cracco2, J V Aranda1.
Abstract
OBJECTIVES: First, to determine the feasibility of an ultra-compact wireless device (microEEG) to obtain multichannel electroencephalographic (EEG) recording in the Neonatal Intensive Care Unit (NICU). Second, to identify problem areas in order to improve wireless EEG performance. STUDYEntities:
Keywords: Apnea; electroencephalography; premature infants; seizure
Mesh:
Year: 2016 PMID: 28009337 PMCID: PMC5181680 DOI: 10.3233/NPM-161643
Source DB: PubMed Journal: J Neonatal Perinatal Med ISSN: 1878-4429
Fig.1The microEEG EEG device.
Fig.2Preterm infant undergoing wireless EEG recording. The recording equipment is contained entirely within the isolette. The device transmits EEG recordings wirelessly to a bedside computer in real time. 1) Wireless EEG recording device 2) Connector cable 3) Electrode cap.
Fig.3Flow diagram of the EEG recording process.
Fig.4Wireless EEG recording in a 30 week corrected gestational age preterm infant displayed using Persyst® software.
Subject characteristics
| Characteristic | Mean (SD) or |
| Maternal age yr | 29.22 (6.4)a |
| Gestational age wk | 27.4 (1.5)a |
| Birth weight g | 988 (222)a |
| Male | 14 (50)b |
| Multiple gestation | 4 (14)b |
| Cesarean delivery | 25 (89)b |
| Antenatal steroids | 24 (85)b |
| 5-min Apgar score <5 | 1 (4)b |
aMean (SD); bn (%).
Comparison of interpretable vs. uninterpretable recordings
| Characteristics at time of recording | All | Interpretable | Uninterpretable | |
| 83a (100) | 65 (78) | 18 (22) | ||
| Corrected gestational age wk, mean (95% CI) | 33.2 (32.6–33.9) | 32.6 (31.9–33.3) | 35.5 (34.2–36.8) | |
| Weight g, mean (95% CI) | 1725 (1602–1847) | 1588 (1444–1732) | 2204 (1952–2455) | |
| Head circumference cm, mean (95% CI) | 28.4 (27.8–28.9) | 27.8 (27.1–28.5) | 30.2 (29.0–31.4) | |
| Chronological age d, mean (95% CI) | 42.4 (37.9–46.9) | 37.1 (32.6–41.7) | 61.4 (52.9–70.0) | |
| Mode of ventilation, | ||||
| Room air | 31 (37) | 20 (31) | 11 (61) | |
| Oxyhood | 1 (1) | 1 (2) | 0 (0) | |
| Nasal cannula | 24 (29) | 21 (32) | 3 (17) | |
| Continuous positive airway pressure | 11 (13) | 10 (15) | 1 (6) | |
| Nasal intermittent mandatory ventilation | 6 (7) | 6 (9) | 0 (0) | |
| Intubated | 10 (12) | 7 (11) | 3 (17) | |
| Method of feeding, | ||||
| No feeding | 9 (10) | 9 (13) | 0 (0) | |
| Orogastric or nasogastric tube | 45 (54) | 38 (58) | 7 (39) | |
| Oral | 29 (35) | 18 (28) | 11 (61) |
aExcludes one recording that could not be accessed. bCalculated using mixed linear model.