Comparative dose levels between CT-scanner and slot-scanning device (EOS system) in pregnant women pelvimetry.

A Ben Abdennebi1, S Aubry2, L Ounalli3, M S Fayache4, E Delabrousse2, Y Petegnief5.   

Abstract

PURPOSE: To estimate fetal absorbed doses for pregnant women pelvimetry, a comparative study between EOS imaging system and low-dose spiral CT-scanner was carried out. For this purpose three different studies were investigated: in vivo, in vitro and Monte Carlo calculations.
METHODS: In vivo dosimetry was performed, using OSL NanoDot dosimeters, to determine the dose to the skin of twenty pregnant women. In vitro studies were established by using a cubic phantom of water, in order to estimate the out of field doses. In the latter study, OSLDs were placed at depths corresponding to the lowest, average and highest position of the uterus. Monte Carlo calculations of effective doses to high radio-sensitive organs were established, using PCXMC and CTExpo software suites for EOS imaging system and CT-scanner, respectively.
RESULTS: The EOS imaging system reduces radiation exposure 4 to 8 times compared to the CT-scanner. The entrance skin doses were 74% (p-values <0.01) higher with the CT-scanner than with the EOS system. In the out of field region, the measured doses of the EOS system were reduced by 80% (p-values <0.02). Monte Carlo calculations confirmed that effective doses to organs are less accentuated for EOS than for CT pelvimetry.
CONCLUSIONS: The EOS system is less irradiating than the CT exam. The out-of-field dose which is significant, is lower in the EOS than in the CT-scanner and could be reduced even further by optimizing the time used for image acquisition.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  CT-scanner; EOS imaging system; Monte Carlo; OSLD NanoDot

Mesh:

Year:  2016        PMID: 27993442     DOI: 10.1016/j.ejmp.2016.12.008

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Asymmetric pelvis and term breech presentation: is planned vaginal delivery still a safe option?

Authors:  M Lallemant; T Ceri; Y Offringa; S Aubry; A Bourtembourg; D Riethmuller; R Ramanah; P Guerby; N Mottet
Journal:  Arch Gynecol Obstet       Date:  2021-03-31       Impact factor: 2.344

2.  Assessing component orientation of total hip arthroplasty using the low-dose bi-planar radiographs.

Authors:  Zhuyi Ma; Hao Tang; Yixin Zhou; Siyuan Wang; Dejin Yang; Shaoyi Guo
Journal:  BMC Musculoskelet Disord       Date:  2022-09-26       Impact factor: 2.562

3.  Equivalent Dose and Risk of Exposure Induced Cancer Death of Different Organs due to Various Image Techniques of EOS Imaging System.

Authors:  Nima Hamzian; Saeid Afereydoon; Mahdi Ghorbani; Seyed Mohammad Jalil Abrisham; Zahra Roozmand; Sepideh Abdollahi-Dehkordi; Morteza Sepehr Javan; Mohammad Reza Deevband
Journal:  J Biomed Phys Eng       Date:  2021-06-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.