| Literature DB >> 2798879 |
E C Hansen1, M C Hawkins, I F Hawkins, E W Akins, S G Miles, J P Collela.
Abstract
The authors describe a new angiographic catheter, which delivers equal flow rates through the side holes and reduced flow through the end hole compared with conventional catheters. Computer analysis of catheter-flow models revealed that placement of larger side holes proximally and smaller side holes distally produces more uniform flow rates out of the holes. This decreases the risk of potentially hazardous delivery of high-pressure jets of contrast material from the end hole. The flow pattern is so uniform that it resembles a cloud of contrast material on injection. This catheter design is particularly applicable for 3-5-F catheters, which require high-pressure injections. Clinical testing in 50 cases revealed no extravasations or unintentional selective injections with use of this catheter.Mesh:
Substances:
Year: 1989 PMID: 2798879 DOI: 10.1148/radiology.173.2.2798879
Source DB: PubMed Journal: Radiology ISSN: 0033-8419 Impact factor: 11.105