Literature DB >> 27956286

Residual astigmatism after toric intraocular lens implantation: Analysis of data from an online toric intraocular lens back-calculator.

Brent A Kramer1, John P Berdahl1, David R Hardten1, Richard Potvin2.   

Abstract

PURPOSE: To evaluate some possible causes for residual astigmatism after toric intraocular lens (IOL) implantation based on an analysis of data from an online toric IOL back-calculator.
DESIGN: Retrospective data review.
METHODS: An online toric back-calculator was designed to allow users to input preoperative toric planning information along with postoperative IOL orientation and refractive results. These were then used to determine the optimum orientation of the IOL to reduce refractive astigmatism. The collected aggregate data were extracted from this calculator to investigate the associated reasons for residual astigmatic refractive error with toric IOLs.
RESULTS: The study analyzed 12 812 records with a mean postoperative refractive astigmatism of 1.89 diopters (D). Refractive astigmatism was significantly higher with higher IOL cylinder power (P < .01) but was not different by IOL manufacturer. Ninety percent of IOLs were not at the ideal orientation, despite 30% being at the preoperative calculated orientation. Misalignment showed a directional bias for some IOLs but not for others. The mean calculated percentage reduction in residual cylinder after reorientation was 50% ± 31% (SD), with the magnitude of residual astigmatism after IOL reorientation expected to be 0.50 D or less in 37% of eyes (4835/12 812). Expected outcomes were significantly different by IOL type.
CONCLUSIONS: Analysis of data from the online toric back-calculator provided insights into the nature of residual astigmatism after toric IOL implantation. The reasons for residual astigmatism in this data set varied by IOL type. FINANCIAL DISCLOSURE: Proprietary or commercial disclosures are listed after the references.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27956286     DOI: 10.1016/j.jcrs.2016.09.017

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Visual outcome and refractive status with monofocal toric intraocular lens implantation to correct astigmatism during cataract surgery.

Authors:  Mayuri Sh Patil; Archana S Nikose; Shadwala Bharti
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

2.  Rotation Characteristics of Three Toric Monofocal Intraocular Lenses.

Authors:  Brent A Kramer; David R Hardten; John P Berdahl
Journal:  Clin Ophthalmol       Date:  2020-12-16

3.  Comparing IOLM700 TK, Berdahl and Hardten astigmatism fix calculator and Barrett Rx formula in managing residual astigmatism due to toric intraocular lens misalignment.

Authors:  Ajay C Sharma; Avani Khetan
Journal:  Indian J Ophthalmol       Date:  2022-02       Impact factor: 1.848

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.