Ali Pourghasem1, Ismail Ebrahimi Takamjani2, Mohammad Taghi Karimi3, Mohammad Kamali4, Mohammad Jannesari5, Iman Salafian6.
Abstract
[Purpose] Standing and walking are impaired in stroke patients. Therefore, assisted devices are required to restore their walking abilities. The ankle foot orthosis with an external powered source is a new type of orthosis. The aim of this study was to evaluate the performance of a powered ankle foot orthosis compared with unpowered orthoses in a stroke patient.Entities:
Keywords: External powered orthosis; Stroke; Walking
Year: 2016 PMID: 27942156 PMCID: PMC5140836 DOI: 10.1589/jpts.28.3236
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Fig. 1.External powered orthosis
The gait parameters of the stroke subject during walking without and with three types of orthosis
| Orthosis type | Test condition | Velocity (m/min) | Stride (m) | Cadence (steps/min) | Force X (N) | Force Y (N) | Force Z (N) |
|---|---|---|---|---|---|---|---|
| Powered AFO | Without orthosis | 9.72 (± 2.85) | 0.46 (± 0.09) | 45.2 (± 6.28) | 33.64 (± 1.77) | 23.1 (± 4.08) | 587.8 (± 16.02) |
| With orthosis | 8.31 (± 0.57) | 0.67 (± 0.039) | 24.76 (± 1.11) | 21.65 (± 14.51) | 27.5 (± 9.94) | 774.66 (± 9.36) | |
| PLS orthosis | Without orthosis | 9.72 (± 2.85) | 0.46 (± 0.09) | 45.2 (± 6.28) | 33.64 (± 1.77) | 23.1 (± 4.08) | 587.8 (± 16.02) |
| With orthosis | 9.12 (± 3.19) | 0.4 (± 0.10) | 42.56 (± 3.16) | 28.16 (± 4.19) | 30.99 (± 3.91) | 638 (± 5.16) | |
| Carbon AFO | Without orthosis | 9.72 (± 2.85) | 0.46 (± 0.09) | 45.2 (± 6.28) | 33.64 (± 1.77) | 23.1 (± 4.08) | 587.8 (± 16.02) |
| With orthosis | 8.46 (± 1.05) | 0.4 (± 0.049) | 42.56 (± 2.32) | 36.78 (± 5.77) | 26.28 (± 1.74) | 627.5 (± 4.04) |
The gait parameters of the stroke subject during walking without and with three types of orthosis
| Orthosis type | Test condition | Ankle joint | Knee joint | Hip joint | Pelvic | ||||
|---|---|---|---|---|---|---|---|---|---|
| Moment (Nm/BM) | Range (degrees) | Moment (Nm/BM) | Range (degrees) | Flexion moment (Nm/BM) | Extension moment (Nm/BM) | Range (degrees) | Range (degrees) | ||
| Powered AFO | Without orthosis | 0.6 (± 0.09) | 22.4 (± 3.11) | 0.133 (± 0.02) | 38.85 (± 2.49) | 0.267 (± 0.03) | 0.554 (± 0.22) | 30.23 (± 3.9) | 15.93 (± 3.12) |
| With orthosis | 0.739 (± 0.11) | 24.5 (± 7.11) | 0.228 (± 0.23) | 38.87 (± 13.6) | 0.396 (± 0.096) | 0.624 (± 0.07) | 27.3 (± 5.90) | 25.66 (± 2.79) | |
| PLS orthosis | Without orthosis | 0.6 (± 0.09) | 22.4 (± 3.11) | 0.133 (± 0.02) | 38.85 (± 2.49) | 0.267 (± 0.03) | 0.554 (± 0.22) | 30.23 (± 3.9) | 15.93 (± 3.12) |
| With orthosis | 0.687 (± 0.18) | 18.13 (± 3.58) | 0.22 (± 0.25) | 31.97 (± 12.1) | 0.421 (± 0.025) | 0.254 (± 0.046) | 29.58 (± 3.89) | 16.18 (± 3.81) | |
| Carbon AFO | Without orthosis | 0.6 (± 0.09) | 22.4 (± 3.11) | 0.133 (± 0.02) | 38.85 (± 2.49) | 0.267 (± 0.03) | 0.554 (± 0.22) | 30.23 (± 3.9) | 15.93 (± 3.12) |
| With orthosis | 0.975 (± 0.04) | 21.8 (± 1.72) | 0.146 (± 0.038) | 39.99 (± 1.35) | 0.407 (± 0.015) | 0.286 (± 0.035) | 39.15 (± 1.75) | 27.24 (± 17.82) | |