Feasibility of Shape-Memory Ni/Ti Alloy Wire Containing Tube Elevators for Transcrestal Detaching Maxillary Sinus Mucosa: Ex Vivo Study.

Yanfeng Li1, Fuli Wang, Pin Hu, Jiadong Fan, Yishi Han, Bin Liu, Tao Liu, Chunhao Yang, Xiangmin Gu.   

Abstract

BACKGROUND: Osteotome sinus floor elevation is a less invasive approach to augment an insufficient alveolar bone at the posterior maxilla for dental implantation. However, this approach has some limitations due to the lack of sinus lift tools available for clinical use and the small transcrestal access to the maxillary sinus floor. We recently invented shape-memory Ni/Ti alloy wire containing tube elevators for transcrestal detaching maxillary sinus mucosa, and developed goat ex vivo models for direct visualizing the effectiveness of detaching sinus mucosa in real time during transcrestal maxillary sinus floor elevation.
METHODS: We evaluated our invented elevators, namely elevator 012 and elevator 014, for their effectiveness for transcrestal detaching maxillary sinus mucosa using the goat ex vivo models. We measured the length of sinus mucosa detached in mesial and distal directions or buccal and palatal directions, and the space volume created by detaching maxillary sinus mucosa in mesial, distal, buccal and palatal directions using the invented elevators.
RESULTS: Elevator 012 had a shape-memory Ni/Ti alloy wire with a diameter of 0.012 inch, while elevator 014 had its shape-memory Ni/Ti alloy wire with a diameter of 0.014 inch. Elevator 012 could detach the goat maxillary sinus mucosa in the mesial or distal direction for 12.1±4.3 mm, while in the buccal or palatal direction for 12.5±6.7 mm. The elevator 014 could detach the goat maxillary sinus mucosa for 23.0±4.9 mm in the mesial or distal direction, and for 19.0±8.1 mm in the buccal or palatal direction. An average space volume of 1.7936±0.2079 ml was created after detaching the goat maxillay sinus mucosa in both mesial/distal direction and buccal/palatal direction using elevator 012; while the average space volume created using elevator 014 was 1.8764±0.2366 ml.
CONCLUSION: Both two newly invented tube elevators could effectively detach the maxillary sinus mucosa on the goat ex vivo sinus models. Moreover, elevator 014 has advantages over the elevator 012 for the capability to detach sinus mucosa.
© 2016 The Author(s) Published by S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27941347     DOI: 10.1159/000453152

Source DB:  PubMed          Journal:  Cell Physiol Biochem        ISSN: 1015-8987


  1 in total

1.  Comparison of biological and mechanical properties of different paranasal sinus mucosa in goat.

Authors:  Xinyu Fang; Cai Mi; Yingdi Wang; Yidan Sun; Jia Nie; Haoning Tang; Yajing Liu; Yanfeng Li; Jing Wang; Zheng Ma; Yishi Han
Journal:  BMC Oral Health       Date:  2022-05-25       Impact factor: 3.747

  1 in total