Literature DB >> 27934777

Rodent model for assessing the long term safety and performance of peripheral nerve recording electrodes.

Srikanth Vasudevan1, Kunal Patel, Cristin Welle.   

Abstract

OBJECTIVE: In the US alone, there are approximately 185 000 cases of limb amputation annually, which can reduce the quality of life for those individuals. Current prosthesis technology could be improved by access to signals from the nervous system for intuitive prosthesis control. After amputation, residual peripheral nerves continue to convey motor signals and electrical stimulation of these nerves can elicit sensory percepts. However, current technology for extracting information directly from peripheral nerves has limited chronic reliability, and novel approaches must be vetted to ensure safe long-term use. The present study aims to optimize methods to establish a test platform using rodent model to assess the long term safety and performance of electrode interfaces implanted in the peripheral nerves. APPROACH: Floating Microelectrode Arrays (FMA, Microprobes for Life Sciences) were implanted into the rodent sciatic nerve. Weekly in vivo recordings and impedance measurements were performed in animals to assess performance and physical integrity of electrodes. Motor (walking track analysis) and sensory (Von Frey) function tests were used to assess change in nerve function due to the implant. Following the terminal recording session, the nerve was explanted and the health of axons, myelin and surrounding tissues were assessed using immunohistochemistry (IHC). The explanted electrodes were visualized under high magnification using scanning electrode microscopy (SEM) to observe any physical damage. MAIN
RESULTS: Recordings of axonal action potentials demonstrated notable session-to-session variability. Impedance of the electrodes increased upon implantation and displayed relative stability until electrode failure. Initial deficits in motor function recovered by 2 weeks, while sensory deficits persisted through 6 weeks of assessment. The primary cause of failure was identified as lead wire breakage in all of animals. IHC indicated myelinated and unmyelinated axons near the implanted electrode shanks, along with dense cellular accumulations near the implant site. Scanning electron microscopy (SEM) showed alterations of the electrode insulation and deformation of electrode shanks. SIGNIFICANCE: We describe a comprehensive testing platform with applicability to electrodes that record from the peripheral nerves. This study assesses the long term safety and performance of electrodes in the peripheral nerves using a rodent model. Under this animal test platform, FMA electrodes record single unit action potentials but have limited chronic reliability due to structural weaknesses. Future work will apply these methods to other commercially-available and novel peripheral electrode technologies.

Entities:  

Mesh:

Year:  2016        PMID: 27934777      PMCID: PMC5335874          DOI: 10.1088/1741-2552/14/1/016008

Source DB:  PubMed          Journal:  J Neural Eng        ISSN: 1741-2552            Impact factor:   5.379


  64 in total

1.  Residual function in peripheral nerve stumps of amputees: implications for neural control of artificial limbs.

Authors:  Gurpreet S Dhillon; Stephen M Lawrence; Douglas T Hutchinson; Kenneth W Horch
Journal:  J Hand Surg Am       Date:  2004-07       Impact factor: 2.230

2.  Double nerve intraneural interface implant on a human amputee for robotic hand control.

Authors:  Paolo M Rossini; Silvestro Micera; Antonella Benvenuto; Jacopo Carpaneto; Giuseppe Cavallo; Luca Citi; Christian Cipriani; Luca Denaro; Vincenzo Denaro; Giovanni Di Pino; Florinda Ferreri; Eugenio Guglielmelli; Klaus-Peter Hoffmann; Stanisa Raspopovic; Jacopo Rigosa; Luca Rossini; Mario Tombini; Paolo Dario
Journal:  Clin Neurophysiol       Date:  2010-01-27       Impact factor: 3.708

Review 3.  Upper limb prosthesis use and abandonment: a survey of the last 25 years.

Authors:  Elaine A Biddiss; Tom T Chau
Journal:  Prosthet Orthot Int       Date:  2007-09       Impact factor: 1.895

4.  Paw withdrawal threshold in the von Frey hair test is influenced by the surface on which the rat stands.

Authors:  G M Pitcher; J Ritchie; J L Henry
Journal:  J Neurosci Methods       Date:  1999-03-01       Impact factor: 2.390

5.  The foreign body response to the Utah Slant Electrode Array in the cat sciatic nerve.

Authors:  M B Christensen; S M Pearce; N M Ledbetter; D J Warren; G A Clark; P A Tresco
Journal:  Acta Biomater       Date:  2014-07-17       Impact factor: 8.947

6.  Anti-TNF-alpha reduces the inflammatory reaction associated with cuff electrode implantation around the sciatic nerve.

Authors:  Valérie Vince; Mårten E Brelen; Jean Delbeke; Ides M Colin
Journal:  J Neuroimmunol       Date:  2005-08       Impact factor: 3.478

Review 7.  Providing a sense of touch to prosthetic hands.

Authors:  Bao Tram Nghiem; Ian C Sando; R Brent Gillespie; Bryan L McLaughlin; Gregory J Gerling; Nicholas B Langhals; Melanie G Urbanchek; Paul S Cederna
Journal:  Plast Reconstr Surg       Date:  2015-06       Impact factor: 4.730

8.  Biocompatibility of platinum-metallized silicone rubber: in vivo and in vitro evaluation.

Authors:  V Vince; M A Thil; C Veraart; I M Colin; J Delbeke
Journal:  J Biomater Sci Polym Ed       Date:  2004       Impact factor: 3.517

9.  Chronic response of the rat sciatic nerve to the flat interface nerve electrode.

Authors:  Dustin J Tyler; Dominique M Durand
Journal:  Ann Biomed Eng       Date:  2003-06       Impact factor: 3.934

10.  Microelectrode array recordings from the ventral roots in chronically implanted cats.

Authors:  Shubham Debnath; Matthew J Bauman; Lee E Fisher; Douglas J Weber; Robert A Gaunt
Journal:  Front Neurol       Date:  2014-07-07       Impact factor: 4.003

View more
  11 in total

1.  In Vitro/Ex Vivo Investigation of Modified Utah Electrode Array to Selectively Sense and Pace the Sub-Surface Cardiac His Bundle.

Authors:  Ankur R Shah; Muhammad S Khan; Annie M Hirahara; Matthias Lange; Ravi Ranjan; Derek J Dosdall
Journal:  ACS Biomater Sci Eng       Date:  2020-05-07

2.  In vitro multichannel single-unit recordings of action potentials from mouse sciatic nerve.

Authors:  L Chen; S J Ilham; T Guo; S Emadi; B Feng
Journal:  Biomed Phys Eng Express       Date:  2017-07-26

3.  The development of neural stimulators: a review of preclinical safety and efficacy studies.

Authors:  Robert K Shepherd; Joel Villalobos; Owen Burns; David A X Nayagam
Journal:  J Neural Eng       Date:  2018-05-14       Impact factor: 5.379

4.  Electrochemistry of a Robust Neural Interface.

Authors:  Pavel A Takmakov
Journal:  Electrochem Soc Interface       Date:  2017

Review 5.  Toward higher-performance bionic limbs for wider clinical use.

Authors:  Dario Farina; Ivan Vujaklija; Rickard Brånemark; Anthony M J Bull; Hans Dietl; Bernhard Graimann; Levi J Hargrove; Klaus-Peter Hoffmann; He Helen Huang; Thorvaldur Ingvarsson; Hilmar Bragi Janusson; Kristleifur Kristjánsson; Todd Kuiken; Silvestro Micera; Thomas Stieglitz; Agnes Sturma; Dustin Tyler; Richard F Ff Weir; Oskar C Aszmann
Journal:  Nat Biomed Eng       Date:  2021-05-31       Impact factor: 25.671

6.  Influence of nerve cuff channel count and implantation site on the separability of afferent ENG.

Authors:  Carolina Silveira; Emma Brunton; Sally Spendiff; Kianoush Nazarpour
Journal:  J Neural Eng       Date:  2018-04-09       Impact factor: 5.379

7.  Characterizing Longitudinal Changes in the Impedance Spectra of In-Vivo Peripheral Nerve Electrodes.

Authors:  Malgorzata M Straka; Benjamin Shafer; Srikanth Vasudevan; Cristin Welle; Loren Rieth
Journal:  Micromachines (Basel)       Date:  2018-11-12       Impact factor: 2.891

8.  Toward optical coherence tomography angiography-based biomarkers to assess the safety of peripheral nerve electrostimulation.

Authors:  Srikanth Vasudevan; Jesse Vo; Benjamin Shafer; Ahhyun S Nam; Benjamin J Vakoc; Daniel X Hammer
Journal:  J Neural Eng       Date:  2019-03-27       Impact factor: 5.379

9.  Classification of naturally evoked compound action potentials in peripheral nerve spatiotemporal recordings.

Authors:  Ryan G L Koh; Adrian I Nachman; José Zariffa
Journal:  Sci Rep       Date:  2019-07-31       Impact factor: 4.379

10.  Regenerative peripheral nerve interfaces for real-time, proportional control of a Neuroprosthetic hand.

Authors:  Christopher M Frost; Daniel C Ursu; Shane M Flattery; Andrej Nedic; Cheryl A Hassett; Jana D Moon; Patrick J Buchanan; R Brent Gillespie; Theodore A Kung; Stephen W P Kemp; Paul S Cederna; Melanie G Urbanchek
Journal:  J Neuroeng Rehabil       Date:  2018-11-20       Impact factor: 4.262

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.