STUDY
DESIGN: Retrospective review of a prospectively collected patient cohort.
OBJECTIVES: This study examined the rate of failure of various types of spinal rod connectors in distraction-based growing rod constructs for early-onset scoliosis. SUMMARY OF BACKGROUND DATA: Distraction-based growing rods have a high rate of reported complications.
METHODS: We performed a retrospective review of a single surgeon's consecutive cases with growing rod constructs for early-onset scoliosis. Connector failure was defined as loosening of the connector-rod interface, which resulted in shortening of the construct and loss of distractive force. We included 30 patients with average age of 5 years (range, 1-10 years) with diagnoses of congenital scoliosis (15), neuromuscular (10), idiopathic (4), and other (1). Minimum follow-up was 2 years, with a mean follow-up of 49 months (range, 24-83 months).
RESULTS: Of 98 connectors, there were 12 connector failures (12.2%) in 9 patients. In all failures, the set screw loosened, resulting in rod slippage. On average, failures occurred 25.5 months after the index procedure, and after the third lengthening. Simple side-to-side closed connectors with circular slots had the highest rate of slippage (7 of 17; 41%). Side-to-side connectors with V-groove slots that mechanically lock the rod had 10 times fewer failures, (1 of 27; 4%) (p = .006). Only 1 of 34 longitudinal connectors failed (3%) (p = .002). The average T1-S1 gain during growth was 6.5 mm/year for patients with a connector failure and 9.0 mm/year for those who did not (p = .25). Connectors failed in dual rods (6 of 18; 33%) and single rods (6 of 15; 40%) at similar rates (p = .97).
CONCLUSIONS: Connector failure in distraction-based growing rods is associated with 28% less T1-S1 gain. Side-to-side connectors with circular slots had a failure rate of 41% compared with side-to-side connectors with V-groove slots (4%). Longitudinal connectors had a failure rate of 4%.
Copyright © 2013. Published by Elsevier Inc.

Entities:  

Keywords:  Connectors; Early-onset scoliosis; Growing rods

Year:  2013        PMID: 27927432     DOI: 10.1016/j.jspd.2012.11.002

Source DB:  PubMed          Journal:  Spine Deform        ISSN: 2212-134X


  2 in total

Review 1.  Current benchtop protocols are not appropriate for the evaluation of distraction-based growing rods: a literature review to justify a new protocol and its development.

Authors:  Niloufar Shekouhi; Amey Kelkar; David Dick; Vijay K Goel; Derek Shaw
Journal:  Eur Spine J       Date:  2022-01-29       Impact factor: 3.134

2.  Clinically relevant finite element technique based protocol to evaluate growing rods for early onset scoliosis correction.

Authors:  Niloufar Shekouhi; David Dick; Maxwell William Baechle; Dilpreet Kaur Kaeley; Vijay K Goel; Hassan Serhan; Jeremy Rawlinson; Derek Shaw
Journal:  JOR Spine       Date:  2020-08-21