| Literature DB >> 27921998 |
M F Orihuela1, A M Somoza, J Colchero, M Ortuño, E Palacios-Lidón.
Abstract
In this work, we propose an intuitive and easily implementable approach to model and interpret scanning Kelvin probe microscopy images of insulating samples with localized charges. The method, based on the image charges method, has been validated by a systematic comparison of its predictions with experimental measurements performed on charge domains of different sizes, injected in polymethyl methacrylate discontinuous films. The agreement between predictions and experimental lateral profiles, as well as with spectroscopy tip-sample distance curves, supports its consistency. The proposed procedure allows obtaining quantitative information such as total charge and the size of a charge domain and allows estimating the most adequate measurement parameters.Entities:
Year: 2016 PMID: 27921998 DOI: 10.1088/1361-6528/28/2/025703
Source DB: PubMed Journal: Nanotechnology ISSN: 0957-4484 Impact factor: 3.874