Literature DB >> 27919511

Metaphyseal locking plate as an external fixator for open tibial fracture: Clinical outcomes and biomechanical assessment.

Ching-Hou Ma1, Chin-Hsien Wu1, Jiun-Ru Jiang2, Yuan-Kun Tu1, Ting-Sheng Lin3.   

Abstract

OBJECTIVE: This study aimed to evaluate the outcome of using a metaphyseal locking plate as a definitive external fixator for treating open tibial fractures based on biomechanical experiments and analysis of clinical results.
METHODS: A metaphyseal locking plate was used as an external fixator in 54 open tibial fractures in 52 patients. The mean follow-up was 38 months (range, 20-52 months). Moreover, static axial compression and torsional tests were performed to evaluate the strength of the fixation techniques.
RESULTS: The average fracture healing time was 34.5 weeks (range, 12-78 weeks). At 4 weeks postoperatively and at the final follow-up, the average Hospital for Special Surgery knee score was 85 (range, 81-100) and 94 (range, 88-100), respectively, and the American Orthopaedic Foot and Ankle Society score was 88 (range, 80-100) and 96 (range, 90-100), respectively. Based on the static test result, the axial stiffness was significantly different among groups (p=0.002), whereas the torsional stiffness showed no significant difference (p=0.068).
CONCLUSIONS: Clinical outcomes show that the use of locking plate as a definitive external fixator is an alternative choice for tibial fractures after obtaining appropriate fracture reduction. However, external locked plating constructs were not as strong as standard locked plating constructs. Therefore, the use of external locked plating constructs as a definitive treatment warrants further biomechanical study for construct strength improvement.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Biomechanical assessment; External fixator; Locking plate; Open tibial fracture

Mesh:

Year:  2016        PMID: 27919511     DOI: 10.1016/j.injury.2016.11.031

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  7 in total

1.  [Treatment of Gustilo-Anderson B type injury of distal femur complicated with bone defect by membrane induction technique].

Authors:  Jinzhu Fan; Fei Cong; Xiaoyu Ren; Ke Zhang; Tao Song
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2022-05-15

2.  Investigating the biomechanical function of the plate-type external fixator in the treatment of tibial fractures: a biomechanical study.

Authors:  Di Shi; Kaiyuan Liu; Haomeng Zhang; Xinli Wang; Guochen Li; Lianhe Zheng
Journal:  BMC Musculoskelet Disord       Date:  2020-02-27       Impact factor: 2.362

3.  Comparison between external locking plate fixation and conventional external fixation for extraarticular proximal tibial fractures: a finite element analysis.

Authors:  Dejan Blažević; Janoš Kodvanj; Petra Adamović; Dinko Vidović; Zlatko Trobonjača; Srećko Sabalić
Journal:  J Orthop Surg Res       Date:  2022-01-11       Impact factor: 2.359

4.  Effectiveness and Biomechanical Analysis of the Ortho-Bridge System on Femoral Fracture Healing: A Retrospective Analysis.

Authors:  Yubin Qi; Lin Yao; Yuntao Long; Guilai Zuo; Qingjie Zhang; Zhenlin Liu; Wen Wang
Journal:  Indian J Orthop       Date:  2022-08-09       Impact factor: 1.033

5.  External Locking Plate Fixation for Femoral Subtrochanteric Fractures.

Authors:  Yen-Chun Chiu; Chin-Hsien Wu; Kun-Ling Tsai; I-Ming Jou; Yuan-Kun Tu; Ching-Hou Ma
Journal:  Geriatr Orthop Surg Rehabil       Date:  2022-09-02

6.  Unilateral External Fixator Combined with Lateral Auxiliary Frame for Ultimate Treatment of Tibia and Fibula Shaft Fractures with Poor Soft Tissue Conditions.

Authors:  Xinhui Wang; Bao Wang; Xizhi Hou; Xiaodong Cheng; Tao Zhang
Journal:  Biomed Res Int       Date:  2022-08-05       Impact factor: 3.246

Review 7.  Locked plating as an external fixator in treating tibial fractures: A PRISMA-compliant systematic review.

Authors:  Peng Luo; Ding Xu; Jia Wu; Yi-Heng Chen
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.817

  7 in total