| Literature DB >> 27918484 |
Yue Deng1,2, Cheng Chen3,4, Xiaojun Xian5, Francis Tsow6,7, Gaurav Verma8, Rob McConnell9, Scott Fruin10, Nongjian Tao11,12, Erica S Forzani13,14.
Abstract
A novel portable wireless volatile organic compound (VOC) monitoring device with disposable sensors is presented. The device is miniaturized, light, easy-to-use, and cost-effective. Different field tests have been carried out to identify the operational, analytical, and functional performance of the device and its sensors. The device was compared to a commercial photo-ionization detector, gas chromatography-mass spectrometry, and carbon monoxide detector. In addition, environmental operational conditions, such as barometric change, temperature change and wind conditions were also tested to evaluate the device performance. The multiple comparisons and tests indicate that the proposed VOC device is adequate to characterize personal exposure in many real-world scenarios and is applicable for personal daily use.Entities:
Keywords: Photo Ionization Detector (PID); epidemiology study; personal monitor; volatile organic compound (VOC)
Mesh:
Substances:
Year: 2016 PMID: 27918484 PMCID: PMC5191041 DOI: 10.3390/s16122060
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1(a) Device dimension and weight; (b) Pictures of sensor components and user interface on a smart phone.
Figure 2Sensor calibration under different concentrations of o-xylene.
Figure 3(a) Scanning sensor QR code; (b) Replacement of new sensor cartridge; (c) Real-time fitness and VOCs exposure device data during personal exercise.
Figure 4Comparison of the response between the new VOC (TVOC) device with a MIP-QTF sensor and RAE Photo-Ionization Detector (PID) for levels assessed during a trip on Los Angeles Highway 101. NOTE: The response of TVOC was calculated in ppmC using a calibration factor for “outdoor environment with motor vehicle exhaust” (see text for more details), while the response of PID was calculated in ppm using the calibration procedure described in the instrument manual.
Figure 5Selectivity validation test to H2S with artificial gas sample (a) and gas sample from Mammoth Spring, Yellowstone National Park. Real-time test was done on the new VOC device (b) and gas sample was collected for GC-MS analysis in the lab (c,d). The single peak in the chromatogram confirmed non-significant concentrations of other VOCs, and a presence of H2S.
Figure 6Outdoor testing of traffic markers as a function of time: (a) Concentration of Carbon Monoxide; (b) Corresponding concentration of total hydrocarbon (HC).
Figure 7The new VOC device’s performance to a rapid elevation increase. Starting point and end point of the trip are indicated using dashed-dotted lines.
Figure 8(a) VOC device’s real-time hydrocarbon concentration test at point A; (b) Corresponding delta temperature between points A and B. Height between point A and B was 14 ft.
Figure 9Wind speed and direction effect on VOC device performance. (a) HC concentration under lower wind speed and wind direction from areas with higher expected concentration (direction A showing in (d), (b) HC concentration under higher wind speed and wind direction from areas with lower expected concentration (direction B showing in (d); (c) peak HC values comparison; (d) a map showing the testing location and wind directions.