| Literature DB >> 27914415 |
Shuzeng Zhang1, Christopher M Kube2, Yongfeng Song3, Xiongbing Li3.
Abstract
A procedure is developed for self-calibration of broadband, spherically focused ultrasonic transducers based on reciprocity. The input and received signals are measured in a pulse-echo configuration. These signals are used in conjunction with a multi-Gaussian beam model to obtain the electromechanical transfer function of the transducer. This calibration procedure is advantageous because it reduces the experimental configuration to a single transducer and a reflector. Experimental results indicate that the transfer function is insensitive to on-axis reflector placement. This result supports the feasibility of integrating the calibration procedure into actual testing in some situations.Year: 2016 PMID: 27914415 DOI: 10.1121/1.4962376
Source DB: PubMed Journal: J Acoust Soc Am ISSN: 0001-4966 Impact factor: 1.840