| Literature DB >> 27910622 |
Teemu Nieminen1, Pasi Lähteenmäki1, Zhenbing Tan1, Daniel Cox1, Pertti J Hakonen1.
Abstract
We present a microwave correlation measurement system based on two low-cost USB-connected software defined radio dongles modified to operate as coherent receivers by using a common local oscillator. Existing software is used to obtain I/Q samples from both dongles simultaneously at a software tunable frequency. To achieve low noise, we introduce an easy low-noise solution for cryogenic amplification at 600-900 MHz based on single discrete HEMT with 21 dB gain and 7 K noise temperature. In addition, we discuss the quantization effects in a digital correlation measurement and determination of optimal integration time by applying Allan deviation analysis.Year: 2016 PMID: 27910622 DOI: 10.1063/1.4966971
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523