Diagnostic accuracy of intraluminal blood speckle intensity on intravascular ultrasound for physiological assessment of coronary artery stenosis.

Yuichi Saito1, Hideki Kitahara, Takashi Nakayama, Yoshihide Fujimoto, Yoshio Kobayashi.   

Abstract

OBJECTIVES: The difference in intraluminal intensity of blood speckle (IBS) on integrated backscatter intravascular ultrasound (IVUS) across the coronary stenosis was reportedly correlated with fractional flow reserve (FFR) in the left descending coronary artery. The aim of this study was to investigate the novel physiological assessment using IVUS in all coronary arteries. PATIENTS AND METHODS: Fifty-four patients with 57 coronary lesions underwent both FFR and IVUS. Intraluminal IBS was analyzed using integrated backscatter IVUS in cross-sections at the ostium and the distal site of the target vessel. ΔIBS was calculated as: (distal IBS)-(ostium IBS).
RESULTS: Both ΔIBS (r=-0.50, P<0.01) and minimum lumen area (MLA) (r=0.55, P<0.01) showed significant correlations with FFR. There were significant correlations between FFR and ΔIBS in the right and left descending coronary arteries (r=-0.60, P=0.02, and r=-0.58, P<0.01), but not in the left circumflex (r=0.30, P=0.44). In receiver operating characteristic curve analyses, ΔIBS predicted FFR less than or equal to 0.80 (area under the curve=0.82, P<0.01, best cutoff value=6.78), as with MLA (area under the curve=0.83, P<0.01, best cutoff value=2.38). FFR progressively decreased in association with ΔIBS greater than or equal to 6.78 and MLA less than or equal to 2.38, and was the lowest when these were combined.
CONCLUSION: ΔIBS was correlated with FFR in right and left descending coronary arteries. IVUS may assess coronary artery stenosis anatomically and physiologically.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 27906702     DOI: 10.1097/MCA.0000000000000457

Source DB:  PubMed          Journal:  Coron Artery Dis        ISSN: 0954-6928            Impact factor:   1.439


  1 in total

1.  Comparison between minimum lumen cross-sectional area and intraluminal ultrasonic intensity analysis using integrated backscatter intravascular ultrasound for prediction of functionally significant coronary artery stenosis.

Authors:  Hironori Takami; Shinjo Sonoda; Yoshitaka Muraoka; Toshiya Miura; Akiyoshi Shimizu; Reo Anai; Yoshinori Sanuki; Tetsu Miyamoto; Yasushi Oginosawa; Yoshihisa Fujino; Yuki Tsuda; Masaru Araki; Yutaka Otsuji
Journal:  Heart Vessels       Date:  2018-07-30       Impact factor: 2.037

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.