Literature DB >> 27906220

Noise analysis of speckle-based x-ray phase-contrast imaging.

Tunhe Zhou, Marie-Christine Zdora, Irene Zanette, Jenny Romell, Hans M Hertz, Anna Burvall.   

Abstract

Speckle-based x-ray phase-contrast imaging has drawn increasing interest in recent years as a simple, multimodal, cost-efficient, and laboratory-source adaptable method. We investigate its noise properties to help further optimization on the method and further comparison with other phase-contrast methods. An analytical model for assessing noise in a differential phase signal is adapted from studies on the digital image correlation technique in experimental mechanics and is supported by simulations and experiments. The model indicates that the noise of the differential phase signal from speckle-based imaging has a behavior similar to that of the grating-based method.

Year:  2016        PMID: 27906220     DOI: 10.1364/OL.41.005490

Source DB:  PubMed          Journal:  Opt Lett        ISSN: 0146-9592            Impact factor:   3.776


  1 in total

1.  Study on the Cytotoxic Microstructure of Titanium Dioxide Nanoparticles by X-Ray Phase-Contrast CT Imaging.

Authors:  Jingtong Fei; Shanshan Nie; Bo Zhang; Xinli Teng; Ying Chen; Yikun Qu; Zhuoxin Cheng; Linqi Guo
Journal:  Contrast Media Mol Imaging       Date:  2022-07-30       Impact factor: 3.009