| Literature DB >> 27899293 |
Alexander A Brescia1, Steven F Bolling2, Himanshu J Patel2.
Abstract
Valvular regurgitation is a recognized complication of aortic and mitral valve procedures, with paravalvular leak the primary mechanism and transvalvular leak less common. The Cor-Knot automated fastener (LSI Solutions, Victor, NY) has been shown to be a safe, viable alternative to knot pushers or manual tying for use in these procedures. Here we present two patients in whom the orientation of the Cor-Knot fastener appeared to result in transvalvular regurgitation necessitating a redo operation. With adjunctive use of Cor-Knot devices, fastener orientation relative to valve geometry must be assessed for optimal results.Entities:
Mesh:
Year: 2016 PMID: 27899293 DOI: 10.1016/j.athoracsur.2016.11.035