Literature DB >> 27885082

THE SIZE-SPECIFIC DOSE ESTIMATE (SSDE) FOR TRUNCATED COMPUTED TOMOGRAPHY IMAGES.

Choirul Anam1,2, Freddy Haryanto2, Rena Widita2, Idam Arif2, Geoff Dougherty3.   

Abstract

The purpose of this study is to investigate truncated axial computed tomography (CT) images in the clinical environment and to produce correction factors for abdomen, thoracic and head regions based on clinical data, in order to accurately predict the water-equivalent diameter (DW) and size-specific dose estimate (SSDE). We investigated axial images of 75 patients who underwent CT examinations. Truncated axial images were characterized by the truncation percentage (TP). Correction factors were calculated by using the value of DW for a certain TP (truncated image) divided by the value of DW for TP = 0% (the non-truncated image). Most of the thorax images acquired for this study were truncated images (86.2%), in the abdomen region about half of the images were truncated (48.1%), and in the head region only a small portion were truncated (9.1%). In the thorax region the value of TP for the truncated images varied up to 50%, in the abdomen region it varied up to 35%, and in the head region it was smaller than 10%. We have shown how to accurately estimate DW and SSDE by applying a correction factor to the truncated images. The correction factors increase exponentially with increasing TP. The corrected DW and SSDE for the truncated images were significant in the thoracic region, but were not significant in the abdomen and head regions.
© The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2017        PMID: 27885082     DOI: 10.1093/rpd/ncw326

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  An Improved Method of Automated Noise Measurement System in CT Images.

Authors:  Choirul Anam; Idam Arif; Freddy Haryanto; Fauzia P Lestari; Rena Widita; Wahyu S Budi; Heri Sutanto; Kusworo Adi; Toshioh Fujibuchi; Geoff Dougherty
Journal:  J Biomed Phys Eng       Date:  2021-04-01

2.  Real-time fully automated dosimetric computation for CT images in the clinical workflow: A feasibility study.

Authors:  Massimiliano Porzio; Choirul Anam
Journal:  Front Oncol       Date:  2022-08-11       Impact factor: 5.738

3.  An algorithm for automated modulation transfer function measurement using an edge of a PMMA phantom: Impact of field of view on spatial resolution of CT images.

Authors:  Choirul Anam; Toshioh Fujibuchi; Wahyu Setia Budi; Freddy Haryanto; Geoff Dougherty
Journal:  J Appl Clin Med Phys       Date:  2018-10-19       Impact factor: 2.102

4.  Comparison of central, peripheral, and weighted size-specific dose in CT.

Authors:  Choirul Anam; Dwi Adhianto; Heri Sutanto; Kusworo Adi; Mohd Hanafi Ali; William Ian Duncombe Rae; Toshioh Fujibuchi; Geoff Dougherty
Journal:  J Xray Sci Technol       Date:  2020       Impact factor: 1.535

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.