Evaluation of different near-infrared spectroscopy technologies for assessment of tissue oxygen saturation during a vascular occlusion test.

Kevin Steenhaut1, Koen Lapage1, Thierry Bové2, Stefan De Hert1, Annelies Moerman3.   

Abstract

An increasing number of NIRS devices are used to provide measurements of peripheral tissue oxygen saturation (StO2). The aim of the present study is to test the hypothesis that despite technological differences between devices, similar trend values will be obtained during a vascular occlusion test. The devices compared are NIRO-200NX, which measures StO2 and oxyhemoglobin by spatially resolved spectroscopy and the Beer-Lambert law, respectively, and INVOS 5100C and Foresight Elite, which both measure StO2 with the Beer-Lambert law, enhanced with the spatial resolution technique. Forty consenting adults scheduled for CABG surgery were recruited. The respective sensors of the three NIRS devices were applied over the brachioradial muscle. Before induction of anesthesia, 3 min of ischemia were induced by inflating a blood pressure cuff at the upper arm, whereafter cuff pressure was rapidly released. Tissue oxygenation measurements included baseline, minimum and maximum values, desaturation and resaturation slopes, and rise time. Comparisons between devices were performed with the Kruskal-Wallis test with post hoc Mann-Whitney pairwise comparisons. Agreement was evaluated using Bland-Altman plots. Oxyhemoglobin measured with NIRO responded faster than the other NIRS technologies to changes in peripheral tissue oxygenation (20 vs. 27-40 s, p ≤ 0.01). When comparing INVOS with Foresight, oxygenation changes were prompter (upslope 311 [92-523]%/min vs. 114[65-199]%/min, p ≤ 0.01) and more pronounced (minimum value 36 [21-48] vs. 45 [40-51]%, p ≤ 0.01) with INVOS. Significant differences in tissue oxygen saturation measurements were observed, both within the same device as between different devices using the same measurement technology.

Entities:  

Keywords:  Near-infrared spectroscopy; Tissue oxygen saturation; Vascular occlusion test

Mesh:

Substances:

Year:  2016        PMID: 27878503     DOI: 10.1007/s10877-016-9962-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  16 in total

1.  Regional tissue oxygen saturation: comparability and reproducibility of different devices.

Authors:  Mirjam Pocivalnik; Gerhard Pichler; Heinz Zotter; Nina Tax; Wilhelm Müller; Berndt Urlesberger
Journal:  J Biomed Opt       Date:  2011-05       Impact factor: 3.170

Review 2.  Near-infrared spectroscopy (NIRS) monitoring in contemporary anesthesia and critical care.

Authors:  A Moerman; P Wouters
Journal:  Acta Anaesthesiol Belg       Date:  2010

3.  Depth sensitivity analysis of functional near-infrared spectroscopy measurement using three-dimensional Monte Carlo modelling-based magnetic resonance imaging.

Authors:  Chemseddine Mansouri; Jean-Pierre L'huillier; Nasser H Kashou; Anne Humeau
Journal:  Lasers Med Sci       Date:  2010-02-09       Impact factor: 3.161

4.  Inconsistent detection of changes in cerebral blood volume by near infrared spectroscopy in standard clinical tests.

Authors:  D Canova; S Roatta; D Bosone; G Micieli
Journal:  J Appl Physiol (1985)       Date:  2011-04-07

5.  Comparison of two devices using near-infrared spectroscopy for the measurement of tissue oxygenation during a vascular occlusion test in healthy volunteers (INVOS® vs. InSpectra™).

Authors:  Ji-Hyun Lee; Yong-Hee Park; Hee-Soo Kim; Jin-Tae Kim
Journal:  J Clin Monit Comput       Date:  2014-07-09       Impact factor: 2.502

6.  Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: a comparison between INVOS and EQUANOX.

Authors:  Jean-Luc Fellahi; Géraldine Butin; Marc-Olivier Fischer; Guillaume Zamparini; Jean-Louis Gérard; Jean-Luc Hanouz
Journal:  J Crit Care       Date:  2013-06-13       Impact factor: 3.425

7.  Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm.

Authors:  Simon Hyttel-Sorensen; Line C Sorensen; Joan Riera; Gorm Greisen
Journal:  Biomed Opt Express       Date:  2011-10-06       Impact factor: 3.732

Review 8.  Journal of Clinical Monitoring and Computing 2015 end of year summary: tissue oxygenation and microcirculation.

Authors:  T W L Scheeren
Journal:  J Clin Monit Comput       Date:  2016-02-20       Impact factor: 2.502

9.  Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers.

Authors:  Rick Bezemer; Alexandre Lima; Dean Myers; Eva Klijn; Michal Heger; Peter T Goedhart; Jan Bakker; Can Ince
Journal:  Crit Care       Date:  2009-11-30       Impact factor: 9.097

10.  Influence of cutaneous and muscular circulation on spatially resolved versus standard Beer-Lambert near-infrared spectroscopy.

Authors:  Alessandro Messere; Silvestro Roatta
Journal:  Physiol Rep       Date:  2013-12-05
View more
  6 in total

1.  Continuous monitoring of interstitial tissue oxygen using subcutaneous oxygen microsensors: In vivo characterization in healthy volunteers.

Authors:  Stephen C Kanick; Peter A Schneider; Bruce Klitzman; Natalie A Wisniewski; Kerstin Rebrin
Journal:  Microvasc Res       Date:  2019-02-08       Impact factor: 3.514

2.  Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3.

Authors:  Arnaud Ferraris; Matthias Jacquet-Lagrèze; Jean-Luc Fellahi
Journal:  J Clin Monit Comput       Date:  2017-05-02       Impact factor: 2.502

Review 3.  The use of a vascular occlusion test combined with near-infrared spectroscopy in perioperative care: a systematic review.

Authors:  C K Niezen; D Massari; J J Vos; T W L Scheeren
Journal:  J Clin Monit Comput       Date:  2022-01-04       Impact factor: 1.977

4.  Altered microvascular reactivity assessed by near-infrared spectroscopy after hepato-pancreato-biliary surgery.

Authors:  Kazuyuki Mizunoya; Yasunori Yagi; Yuji Morimoto; Satoshi Hirano
Journal:  J Clin Monit Comput       Date:  2021-04-07       Impact factor: 1.977

5.  Mitochondrial oxygen monitoring with COMET: verification of calibration in man and comparison with vascular occlusion tests in healthy volunteers.

Authors:  R Ubbink; M A Wefers Bettink; W van Weteringen; E G Mik
Journal:  J Clin Monit Comput       Date:  2020-10-21       Impact factor: 1.977

6.  Absolute quantification of cerebral tissue oxygen saturation with multidistance broadband NIRS in newborn brain.

Authors:  Zuzana Kovacsova; Gemma Bale; Subhabrata Mitra; Frédéric Lange; Ilias Tachtsidis
Journal:  Biomed Opt Express       Date:  2021-01-15       Impact factor: 3.732

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.