Literature DB >> 27871844

An LC-MS/MS based candidate reference method for the quantification of total gentamicin in human serum and plasma using NMR characterized calibrator material.

Stephanie Lucha1, Judith Taibon1, Stephan Pongratz1, Christian Geletneky1, Erasmus Huber1, Christine Wintterle-Roehm1, Robert Lang1, Stefanie H Grimm2, Thomas Duelffer1, Kirill Tarasov1, Johannes Zander3, Michael Vogeser4, Uwe Kobold1.   

Abstract

BACKGROUND: Accurate measurement of gentamicin concentration in serum and plasma is required for therapeutic drug monitoring to ensure appropriate treatment of patients. In this work, we present a validated LC-MS/MS-based candidate reference measurement procedure for total gentamicin quantification to be used for standardization and harmonization of routine assays applied for therapeutic drug monitoring of this compound. Total gentamicin is the sum of the concentrations of five known congeners C1, C1a, C2, C2a and C2b. To our knowledge, there is so far no LC-MS method for quantification of total gentamicin in human serum described in literature.
METHODS: Sample preparation was based on sample dilution with an aqueous internal standard solution followed by protein precipitation. Stable derivatives of gentamicin-glycine congeners were prepared by chemical synthesis and used as internal standards. The primary calibration material used in this assay was characterized by NMR spectroscopy and the pattern of the gentamicin congeners was determined. The total gentamicin was reported as the sum of the congeners which were quantified individually by LC-MS/MS.
RESULTS: The method allows the measurement of total gentamicin in human serum and plasma in the concentration range of 0.1 to 12.0μg/ml with an assay imprecision of ≤6% CV and an assay accuracy between 96% and 114%. LOD and LOQ for the total gentamicin were 0.04μg/ml and 0.13μg/ml, respectively. Comparative measurement of 128 native patient samples using this method implemented at two laboratory sites showed an excellent agreement.
CONCLUSIONS: Validation results proved that this protocol describes a robust and reliable method which is suggested as reference measurement procedure for the standardization and harmonization of routine assays for the quantification of total gentamicin.
Copyright © 2016. Published by Elsevier B.V.

Entities:  

Keywords:  Gentamicin; LC-MS/MS; NMR; Plasma; Reference method; Serum

Mesh:

Substances:

Year:  2016        PMID: 27871844     DOI: 10.1016/j.cca.2016.11.029

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

Review 1.  The role of mass spectrometry in antibiotic stewardship.

Authors:  Johannes Zander; Michael Paal; Michael Vogeser
Journal:  Clin Mass Spectrom       Date:  2018-11-16
  1 in total