Literature DB >> 27869855

Analytical and simulation results of a triple micro whispering gallery mode probe system for a 3D blood flow rate sensor.

Prateep Phatharacorn, Surasak Chiangga, Preecha Yupapin.   

Abstract

The whispering gallery mode (WGM) is generated by light propagating within a nonlinear micro-ring resonator, which is modeled and made by an InGaAsP/InP material, and called a Panda ring resonator. An imaging probe can also be formed by the micro-conjugate mirror function for the appropriate Panda ring parameter control. The 3D WGM probe can be generated and used for a 3D sensor head and imaging probe. The analytical details and simulation results are given, in which the simulation results are obtained by using the MATLAB and Optiwave programs. From the obtained results, such a design system can be configured to be a thin-film sensor system that can contact the sample surface for the required measurements The outputs of the system are in the form of a WGM beam, in which the 3D WGM probe is also available with the micro-conjugate mirror function. Such a 3D probe can penetrate into the blood vessel and content, from which the time delay among those probes can be detected and measured, and where finally the blood flow rate can be calculated and the blood content 3D image can also be seen and used for medical diagnosis. The tested results have shown that the blood flow rate of 0.72-1.11  μs<sup>-1</sup>, with the blood density of 1060  kgm<sup>-3</sup>, can be obtained.

Mesh:

Year:  2016        PMID: 27869855     DOI: 10.1364/AO.55.009504

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  Integrating Metamaterial Antenna Node and LiFi for Privacy Preserving Intelligent COVID-19 Hospital Patient Management.

Authors:  A Garhwal; M Bunruangses; A E Arumona; P Youplao; K Ray; S Suwandee; P Yupapin
Journal:  Cognit Comput       Date:  2021-01-11       Impact factor: 5.418

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.