Literature DB >> 27866134

One-year clinical outcomes of patients treated with everolimus-eluting bioresorbable vascular scaffolds versus everolimus-eluting metallic stents: a propensity score comparison of patients enrolled in the ABSORB EXTEND and SPIRIT trials.

José de Ribamar Costa1, Alexandre Abizaid, Antonio L Bartorelli, Robert Whitbourn, Nigel Jepson, Marco Perin, Clemens Steinwender, Marrianne Stuteville, Divine Ediebah, Krishnankutty Sudhir, Patrick W Serruys.   

Abstract

AIMS: We sought to compare the outcomes of low/moderate complexity patients treated with the Absorb BVS from the ABSORB EXTEND trial with patients treated with the XIENCE everolimus-eluting stent (EES), using propensity score (PS) matching of pooled data from the SPIRIT trials (SPIRIT II, SPIRIT III, SPIRIT IV) and the XIENCE V USA trial. METHODS AND
RESULTS: ABSORB EXTEND was a prospective, single-arm, open-label clinical study in which 812 patients were enrolled at 56 sites. This study allowed the treatment of lesions ≤28 mm in length and with a reference vessel diameter of 2.0-3.8 mm (as assessed by online QCA). The propensity score was obtained by fitting a logistic regression model with the cohort indicator as the binary outcome and other variables as the predictor variables. At one-year clinical follow-up, there was no statistical difference between groups with regard to MACE (5.0% vs. 4.8%, p=0.83), target lesion failure (5.0% vs. 4.7%, p=0.74), ischaemia-driven target vessel revascularisation (2.3% vs. 3.0%, p=0.38) and device thrombosis (1.0% vs. 0.3%, p=0.11). Myocardial infarction was higher with Absorb (3.3% vs. 1.5%, p=0.02), at the expense of periprocedural CK-MB elevation. Independent predictors of MACE among patients receiving Absorb BVS were treatment of multivessel disease, insulin-dependent diabetes and performance of post-dilation.
CONCLUSIONS: At one-year follow-up, propensity score-matched analysis demonstrated that the clinical safety and effectiveness of Absorb are comparable to those of XIENCE EES among non-complex patients treated with PCI.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27866134     DOI: 10.4244/EIJV12I10A206

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  1 in total

1.  Safety and efficacy outcomes of 3rd generation DES in an all-comer population of patients undergoing PCI: 12-month and 24-month results of the e-Biomatrix French registry.

Authors:  Eric Maupas; Janusz Lipiecki; Raphy Levy; Benjamin Faurie; Bernard Karsenty; Marc Eric Moulichon; François Brunelle; Luc Maillard; Fabien de Poli; Thierry Lefèvre
Journal:  Catheter Cardiovasc Interv       Date:  2017-05-22       Impact factor: 2.692

  1 in total