Literature DB >> 27864458

Three-Dimensional Analysis of Component Stability of the Nellix Endovascular Aneurysm Sealing System After Treatment of Infrarenal Abdominal Aortic Aneurysms.

Bernhard Dorweiler1, Christian Boedecker1, Friedrich Dünschede1, Christian F Vahl1, Marwan Youssef1.   

Abstract

PURPOSE: To assess short-term stability and conformational changes of the Nellix EndoVascular Aneurysm Sealing (EVAS) System using 3-dimensional (3D) analysis.
METHODS: Postoperative computed tomography (CT) scans obtained at 0, 3, and 12 months in 24 patients (mean age 75±7 years; 22 men) who underwent EVAS between December 2013 and December 2014 for intact abdominal aortic aneurysm (within the instructions for use) were evaluated for stent-graft deviation in multiple planes using dedicated 3D analysis software. In addition, 2D analysis using an anatomically fixed reference landmark was performed to assess craniocaudal migration. Clinical and follow-up data of the patients were recorded and matched with results of the imaging analysis.
RESULTS: Overall stability of the Nellix endografts was promising. Relevant conformational changes in the majority of cases were limited to the iliac graft segment and were clinically benign in all cases. Conversely, the only deviation of the proximal stent-graft segment was found in a patient with type Ia endoleak. Additional 2D analysis found relevant (≥5 mm) caudal migration of the Nellix stent-graft in 6 patients, including the one with the type Ia endoleak. In 3 patients, 3D analysis demonstrated the absence of relevant conformational changes of the endografts despite caudal migration.
CONCLUSION: Overall stability of the separate EVAS stent-grafts is promising in the short term. Relevant conformational changes (stent-graft deviation) in the majority of cases were benign and confined to the iliac segment.

Entities:  

Keywords:  2D/3D analysis; abdominal aortic aneurysm; computed tomography angiography; conformational changes; endograft; endoleak; endovascular aneurysm sealing; migration; stent-graft

Mesh:

Year:  2016        PMID: 27864458     DOI: 10.1177/1526602816678201

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  2 in total

1.  Two-Year Outcomes of the Nellix EndoVascular Aneurysm Sealing System for Treatment of Abdominal Aortic Aneurysms.

Authors:  Aleksandra C Zoethout; Johannes T Boersen; Jan M M Heyligers; Jean-Paul P M de Vries; Clark J A M Zeebregts; Michel M P J Reijnen
Journal:  J Endovasc Ther       Date:  2018-03-29       Impact factor: 3.487

2.  Outcome of Nellix-EVAS: single center mid-term results.

Authors:  Arash Najafi; Gabriel Tobias Sheikh; Pius Wigger; Christoph A Binkert
Journal:  CVIR Endovasc       Date:  2019-04-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.