Literature DB >> 27862338

Efficacy and Safety of the GuideLiner Mother-in-Child Guide Catheter Extension in Percutaneous Coronary Intervention.

Divyesh Sharma1, Ashish Shah1, Mark Osten1, Douglas Ing1, Alan Barolet1, Christopher B Overgaard1, Vladimír Džavík1, Peter H Seidelin1.   

Abstract

BACKGROUND: Percutaneous coronary intervention (PCI) procedures are increasingly performed on complex tortuous and heavily calcified coronary lesions. The GuideLiner® catheter (Vascular Solutions, Inc. Minneapolis, MN) is a rapid exchange guide catheter extension system. There is paucity of data about the efficacy of this device in complex PCI. Therefore, we aim to report the indications, success, and efficacy of GuideLiner use in treating complex lesion subset in the present study.
METHODS: We collected data from all patients in whom the GuideLiner was used to facilitate PCI between April 1, 2011 and December 31, 2014 at a tertiary referral center in Toronto, Canada. Demographic and procedural data were obtained from an institutional prospective data registry. Indications for use of the GuideLiner, and angiographic and procedural data specific to the device were also collected.
RESULTS: The GuideLiner was utilized in 307 (317 lesions) of 6105 PCI procedures performed at our institution during the study period. The mean age was 68.4 ± 11.2 years and 82% were male. The majority of lesions were types B2 or C (98%), calcified (86.3%), with proximal tortuosity (88.6%). The overall technical success rate of the GuideLiner was 88%. The procedural success rate was 98.7%. The complication rate was low (1.6%).
CONCLUSIONS: Use of the GuideLiner catheter facilitated successful completion of PCI procedures in a majority of patients with complex lesions. Operators should be aware of the potential complications associated with GuideLiner use.
© 2016, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2016        PMID: 27862338     DOI: 10.1111/joic.12354

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  2 in total

1.  Patient outcomes in GuideLiner facilitated percutaneous coronary intervention stratified by the SYNTAX score: A retrospective analysis.

Authors:  Shuangbo Liu; Christopher Parr; Hannah Zhang; Basem Elbarouni; Ashish Shah; Malek Kass; Amir Ravandi
Journal:  JRSM Cardiovasc Dis       Date:  2019-03-20

2.  Application of a rapid exchange extension catheter technique in type B2/C nonocclusive coronary intervention via a transradial approach.

Authors:  Hong-Chao Wang; Wei Lu; Zi-Han Gao; Ya-Nan Xie; Jie Hao; Jin-Ming Liu
Journal:  World J Clin Cases       Date:  2021-04-26       Impact factor: 1.337