In vivo skin dose measurement using MOSkin detectors in tangential breast radiotherapy.

W L Jong1, N M Ung2, J H D Wong3, K H Ng3, W Z Wan Ishak1, R Abdul Malik1, V C E Phua1, D L Cutajar4, P E Metcalfe4, A B Rosenfeld4, G F Ho1.   

Abstract

The purpose of this study is to measure patient skin dose in tangential breast radiotherapy. Treatment planning dose calculation algorithm such as Pencil Beam Convolution (PBC) and in vivo dosimetry techniques such as radiochromic film can be used to accurately monitor radiation doses at tissue depths, but they are inaccurate for skin dose measurement. A MOSFET-based (MOSkin) detector was used to measure skin dose in this study. Tangential breast radiotherapies ("bolus" and "no bolus") were simulated on an anthropomorphic phantom and the skin doses were measured. Skin doses were also measured in 13 patients undergoing each of the techniques. In the patient study, the EBT2 measurements and PBC calculation tended to over-estimate the skin dose compared with the MOSkin detector (p<0.05) in the "no bolus radiotherapy". No significant differences were observed in the "bolus radiotherapy" (p>0.05). The results from patients were similar to that of the phantom study. This shows that the EBT2 measurement and PBC calculation, while able to predict accurate doses at tissue depths, are inaccurate in predicting doses at build-up regions. The clinical application of the MOSkin detectors showed that the average total skin doses received by patients were 1662±129cGy (medial) and 1893±199cGy (lateral) during "no bolus radiotherapy". The average total skin doses were 4030±72cGy (medial) and 4004±91cGy (lateral) for "bolus radiotherapy". In some cases, patient skin doses were shown to exceed the dose toxicity level for skin erythema. Hence, a suitable device for in vivo dosimetry is necessary to accurately determine skin dose. Copyright Â
© 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Breast conserving radiotherapy; Chest wall irradiation; In vivo dosimetry; MOSFET; Superficial dose; Surface dose

Mesh:

Year:  2016        PMID: 27842982     DOI: 10.1016/j.ejmp.2016.10.022

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Passive radiofrequency x-ray dosimeter tag based on flexible radiation-sensitive oxide field-effect transistor.

Authors:  Tobias Cramer; Ilaria Fratelli; Pedro Barquinha; Ana Santa; Cristina Fernandes; Franck D'Annunzio; Christophe Loussert; Rodrigo Martins; Elvira Fortunato; Beatrice Fraboni
Journal:  Sci Adv       Date:  2018-06-29       Impact factor: 14.136

2.  Low-intensity ultrasound combined with allogenic adipose-derived mesenchymal stem cells (AdMSCs) in radiation-induced skin injury treatment.

Authors:  Zeinab Hormozi Moghaddam; Manijhe Mokhtari-Dizaji; Mohammad Ali Nilforoshzadeh; Mohsen Bakhshandeh; Sahar Ghaffari Khaligh
Journal:  Sci Rep       Date:  2020-11-17       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.