Intraoperative OCT Imaging of the Argus II Retinal Prosthesis System.

Aleksandra V Rachitskaya, Alex Yuan, Meghan J Marino, Jamie Reese, Justis P Ehlers.   

Abstract

BACKGROUND AND
OBJECTIVE: Optimal placement of the Argus II Retinal Prosthesis System (Second Sight Medical Products, Sylmar, CA) is critical. Intraoperative optical coherence tomography (OCT) allows for intrasurgical visualization and confirmation of array placement. In this study, two different OCT systems were evaluated to assess the feasibility and utility of this technology during Argus II surgery. PATIENTS AND METHODS: Intraoperative OCT was performed on five patients undergoing Argus II implantation at Cole Eye Institute from June 2015 to July 2016. The EnVisu portable OCT (Bioptigen, Morrisville, NC) and microscope-integrated RESCAN 700 (Zeiss, Oberkochen, Germany) intraoperative OCT systems were utilized. The EnVisu was used in three patients and the RESCAN 700 in three of the five patients. Following array tacking, intraoperative OCT was performed over the entire array including the edges and tack.
RESULTS: Intraoperative OCT allowed for visualization of the array/retina interface. Microscope integration of the OCT system facilitated ease of focusing, real-time feedback, surgeon-directed OCT scanning to the areas of interest, and enhanced image quality at points of interest.
CONCLUSIONS: Intraoperative imaging of the Argus II electrode array is feasible and provides information about electrode array-retina interface and distance to help guide a surgeon. Microscope integration of OCT appears to provide an optimal and efficient approach to intraoperative OCT during Argus II array placement. [Ophthalmic Surg Lasers Imaging Retina. 2016;47:999-1003.]. Copyright 2016, SLACK Incorporated.

Entities:  

Mesh:

Year:  2016        PMID: 27842194      PMCID: PMC5546736          DOI: 10.3928/23258160-20161031-03

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging Retina        ISSN: 2325-8160            Impact factor:   1.300


  7 in total

Review 1.  Retinal prostheses: current clinical results and future needs.

Authors:  James D Weiland; Alice K Cho; Mark S Humayun
Journal:  Ophthalmology       Date:  2011-11       Impact factor: 12.079

2.  Pathology of damaging electrical stimulation in the retina.

Authors:  L Colodetti; J D Weiland; S Colodetti; A Ray; M J Seiler; D R Hinton; M S Humayun
Journal:  Exp Eye Res       Date:  2007-03-07       Impact factor: 3.467

3.  Erratum to: Worldwide Argus II implantation: recommendations to optimize patient outcomes.

Authors:  Devon H Ghodasra; Adrienne Chen; J Fernando Arevalo; David G Birch; Kari Branham; Brian Coley; Gislin Dagnelie; Eugene de Juan; Robert G Devenyi; Jessy D Dorn; Andy Fisher; Duane R Geruschat; Ninel Z Gregori; Robert J Greenberg; Paul Hahn; Allen C Ho; Ashley Howson; Suber S Huang; Raymond Iezzi; Naheed Khan; Byron L Lam; Jennifer I Lim; Kirsten G Locke; Michelle Markowitz; Anne-Marie Ripley; Mark Rankin; Hannah Schimitzek; Fay Tripp; James D Weiland; Jiong Yan; David N Zacks; K Thiran Jayasundera
Journal:  BMC Ophthalmol       Date:  2016-07-29       Impact factor: 2.209

4.  Ten-Year Follow-up of a Blind Patient Chronically Implanted with Epiretinal Prosthesis Argus I.

Authors:  Lan Yue; Paulo Falabella; Punita Christopher; Varalakshmi Wuyyuru; Jessy Dorn; Paulo Schor; Robert J Greenberg; James D Weiland; Mark S Humayun
Journal:  Ophthalmology       Date:  2015-09-18       Impact factor: 12.079

5.  Determination of feasibility and utility of microscope-integrated optical coherence tomography during ophthalmic surgery: the DISCOVER Study RESCAN Results.

Authors:  Justis P Ehlers; Jeff Goshe; William J Dupps; Peter K Kaiser; Rishi P Singh; Richard Gans; Jonathan Eisengart; Sunil K Srivastava
Journal:  JAMA Ophthalmol       Date:  2015-10       Impact factor: 7.389

6.  Factors Affecting Perceptual Threshold in Argus II Retinal Prosthesis Subjects.

Authors:  A K Ahuja; J Yeoh; J D Dorn; A Caspi; V Wuyyuru; M J McMahon; M S Humayun; R J Greenberg; L Dacruz
Journal:  Transl Vis Sci Technol       Date:  2013-04-12       Impact factor: 3.283

7.  Argus II retinal prosthesis malrotation and repositioning with intraoperative optical coherence tomography in a posterior staphyloma.

Authors:  Michael I Seider; Paul Hahn
Journal:  Clin Ophthalmol       Date:  2015-11-24
  7 in total
  8 in total

1.  The DISCOVER Study 3-Year Results: Feasibility and Usefulness of Microscope-Integrated Intraoperative OCT during Ophthalmic Surgery.

Authors:  Justis P Ehlers; Yasha S Modi; Paula E Pecen; Jeff Goshe; William J Dupps; Aleksandra Rachitskaya; Sumit Sharma; Alex Yuan; Rishi Singh; Peter K Kaiser; Jamie L Reese; Carmen Calabrise; Allison Watts; Sunil K Srivastava
Journal:  Ophthalmology       Date:  2018-03-02       Impact factor: 12.079

2.  Retinal Anatomy and Electrode Array Position in Retinitis Pigmentosa Patients After Argus II Implantation: An International Study.

Authors:  Ninel Z Gregori; Natalia F Callaway; Catherine Hoeppner; Alex Yuan; Aleksandra Rachitskaya; William Feuer; Hossein Ameri; J Fernando Arevalo; Albert J Augustin; David G Birch; Gislin Dagnelie; Salvatore Grisanti; Janet L Davis; Paul Hahn; James T Handa; Allen C Ho; Suber S Huang; Mark S Humayun; Raymond Iezzi; K Thiran Jayasundera; Gregg T Kokame; Byron L Lam; Jennifer I Lim; Naresh Mandava; Sandra R Montezuma; Lisa Olmos de Koo; Peter Szurman; Lejla Vajzovic; Peter Wiedemann; James Weiland; Jiong Yan; David N Zacks
Journal:  Am J Ophthalmol       Date:  2018-06-27       Impact factor: 5.258

3.  THE INTEGRATIVE SURGICAL THEATER: Combining Intraoperative Optical Coherence Tomography and 3D Digital Visualization for Vitreoretinal Surgery in the DISCOVER Study.

Authors:  Justis P Ehlers; Atsuro Uchida; Sunil K Srivastava
Journal:  Retina       Date:  2018-09       Impact factor: 4.256

4.  Model-Based Recommendations for Optimal Surgical Placement of Epiretinal Implants.

Authors:  Michael Beyeler; Geoffrey M Boynton; Ione Fine; Ariel Rokem
Journal:  Med Image Comput Comput Assist Interv       Date:  2019-10-10

5.  Assessment of Postoperative Morphologic Retinal Changes by Optical Coherence Tomography in Recipients of an Electronic Retinal Prosthesis Implant.

Authors:  Stanislao Rizzo; Laura Cinelli; Lucia Finocchio; Ruggero Tartaro; Francesca Santoro; Ninel Z Gregori
Journal:  JAMA Ophthalmol       Date:  2019-03-01       Impact factor: 7.389

6.  Implantation of multiple suprachoroidal electrode arrays in rabbits.

Authors:  Walid Abdallah; Wen Li; James Weiland; Mark Humayun; Hossein Ameri
Journal:  J Curr Ophthalmol       Date:  2017-12-09

7.  Hints for Gentle Submacular Injection in Non-Human Primates Based on Intraoperative OCT Guidance.

Authors:  Gavin S W Tan; Zengping Liu; Tanja Ilmarinen; Veluchamy A Barathi; Caroline K Chee; Gopal Lingam; Xinyi Su; Boris V Stanzel
Journal:  Transl Vis Sci Technol       Date:  2021-01-07       Impact factor: 3.283

8.  Assessment of the Electronic Retinal Implant Alpha AMS in Restoring Vision to Blind Patients with End-Stage Retinitis Pigmentosa.

Authors:  Thomas L Edwards; Charles L Cottriall; Kanmin Xue; Matthew P Simunovic; James D Ramsden; Eberhart Zrenner; Robert E MacLaren
Journal:  Ophthalmology       Date:  2017-10-27       Impact factor: 12.079