Analysis of plasticizers in PVC medical devices: Performance comparison of eight analytical methods.

L Bernard1, D Bourdeaux2, B Pereira3, N Azaroual4, C Barthélémy4, C Breysse5, P Chennell2, R Cueff6, T Dine4, T Eljezi6, F Feutry4, S Genay4, N Kambia4, M Lecoeur4, M Masse7, P Odou7, T Radaniel4, N Simon7, C Vaccher4, C Verlhac6, M Yessad6, B Décaudin7, V Sautou2.   

Abstract

A wide variety of medical devices (MDs) used in hospitals are made of flexible plasticized polyvinylchloride (PVC). Different plasticizers are present in variable amounts in the PVC matrix of the devices and can leach out into the infused solutions and may enter into contact with the patients. The ARMED1 project aims to assess the migration of these plasticizers from medical devices and therefore the level of exposure in patients. For the first task of the project, eight methods were developed to directly detect and quantify the plasticizers in the PVC matrix of the MDs. We compared the overall performances of the analytical methods using standardized and validated criteria in order to provide the scientific community with the guidance and the technical specifications of each method for the intended application. We have shown that routine rapid screening could be performed directly on the MDs using the FTIR technique, with cost-effective analyses. LC techniques may also be used, but with limits and only with individual quantification of the main plasticizers expected in the PVC matrix. GC techniques, especially GC-MS, are both more specific and more sensitive than other techniques. NMR is a robust and specific technique to precisely discriminate all plasticizers in a MD but is limited by its cost and its low ability to detect and quantify plasticizer contamination, e.g. by DEHP. All these results have been confirmed by a real test, called the " blind test " carried out on 10 MD samples.
Copyright © 2016 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Analytical methods; Comparison; Medical devices; Plasticizers

Year:  2016        PMID: 27837878     DOI: 10.1016/j.talanta.2016.10.033

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  3 in total

1.  Development and Validation of Gas Chromatography-Triple Quadrupole Mass Spectrometric Method for Quantitative Determination of Regulated Plasticizers in Medical Infusion Sets.

Authors:  So Hyeon Jeon; Yong Pyo Kim; Younglim Kho; Jeoung Hwa Shin; Won Hyun Ji; Yun Gyong Ahn
Journal:  J Anal Methods Chem       Date:  2018-02-05       Impact factor: 2.193

2.  Compact NMR Spectroscopy for Low-Cost Identification and Quantification of PVC Plasticizers.

Authors:  Anton Duchowny; Alina Adams
Journal:  Molecules       Date:  2021-02-25       Impact factor: 4.411

3.  Augmentation of curved tip of left-sided double-lumen tubes to reduce right bronchial misplacement: A randomized controlled trial.

Authors:  Jeong-Hwa Seo; Susie Yoon; Se-Hee Min; Hyung Sang Row; Jae-Hyon Bahk
Journal:  PLoS One       Date:  2019-01-15       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.