Literature DB >> 27833505

Another utility of signature capnogram: Detection of sodalime canister detachment from anesthesia machine.

Sanjay Kumar1, Manjaree Mishra2, Amrita Bhattacharyya1.   

Abstract

Entities:  

Year:  2016        PMID: 27833505      PMCID: PMC5044746          DOI: 10.4103/1658-354X.177336

Source DB:  PubMed          Journal:  Saudi J Anaesth


× No keyword cloud information.
Sir, Capnography is the continuous carbon dioxide (CO2) level monitoring along with its graphical representation during the respiratory cycle. CO2 rebreathing during general anesthesia is defined as an increased inspired pressure of carbon dioxide above 2 mmHg.[1] The “Signature capnogram” is a type of rebreathing capnogram during the inspiratory phase of the respiratory cycle. It has been already reported in the case of rebreathing with Bain circuit assisted positive pressure ventilation and totally exhausted CO2 absorbent.[23] We like to report the similar signature capnogram during total disconnection of single chambered absorbent canister from the anesthesia machine Datex ohmeda S/5 Avance (Datex-Ohmeda, P.O. Box 7550, Madison WI, 53707-7550, USA). A 30-year-old, 50 kg female with fibroadenoma of right breast was posted for excision. Patient was induced with fentanyl 3 µg/kg and propofol 2 mg/kg after starting standard anesthesia monitoring on Datex ohmeda S/5 Avance anesthesia machine. Orotracheal intubation was facilitated by vecuronium 0.08 mg/kg and mechanical ventilation started. Anesthesia was maintained with 50% air-oxygen mixture and propofol infusion at the rate of 100-200 µg/kg/min. The patient was hemodynamically stable. Capnography was under normal range with EtCO2 value of 36 mmHg and zero inspiratory level. In our center, we do change soda lime at a regular interval. This time anesthesia technician came to replace the exhausted CO2 absorbent with the fresh one during this operation. We took the patient on manual positive pressure ventilation during changing of soda lime. Technician removed the absorbent canister after detaching it. He reattached the canister after putting fresh soda lime in it [Figure 1]. Capnography was normal after attaching the canister for 10 min, but gradually EtCO2 started increasing with raised inspiratory level with an inspiratory peak. The expiratory and inspiratory valves were changed, but abnormal capnogram was still persisting. Rebreathing continued to be apparent. Meanwhile, we found that absorbent canister containing fresh soda lime was hanging from the anesthesia machine making an angle of 45° with the horizontal axis thereby inlet and outlet point of canister were not connected to corresponding outlet and inlet in the machine [Figure 2a]. After proper attachment, this abnormal (signature) capnogram disappeared to normal capnogram after 3 min. Possibilities are that canister and machine could not be attached properly or loosely attached to each other but their inlet and outlet were enough connected to allow gas passage through CO2 absorbent to absorb CO2 giving normal capnogram. After 10 min canister could have detached totally at loose end from the machine but other end was still attached from the machine at 45° from horizontal. In this position, inlet and outlet of the canister were so far from corresponding outlet and inlet of the machine causing automatic closure of valve at the corresponding point in the machine resulting in bypass of exhaled CO2 enriched gas. This exhaled CO2 rich gas collects at the patient end of the inspiratory limb and predominantly fresh gas (gases collected during expiratory pause resulting in further decrease of CO2 concentration) collects at the machine end of the inspiratory limb to produce relative differences in the CO2 concentration in the inspiratory limb during inspiration. This inspiratory CO2 differences results in inspiratory peak followed by an inspiratory dip in CO2 level producing a signature wave in the inspiratory phase of capnography [Figure 2b].
Figure 1

Properly attached single chambered sodalime canister with Anesthesia machine (Datex ohmeda S/5 Avance machine) at both ends leaving no gap between them

Figure 2a

Canister hanging at 45° from hinge point because of detachment from other end and disconnection between inlet and outlet of canister and the machine

Figure 2b

Raised inspiratory fraction of CO2 with inspiratory peak (signature capnogram with rebreathing) resulted from inhalation of CO2 rich gas during the positive pressure ventilation

Properly attached single chambered sodalime canister with Anesthesia machine (Datex ohmeda S/5 Avance machine) at both ends leaving no gap between them Canister hanging at 45° from hinge point because of detachment from other end and disconnection between inlet and outlet of canister and the machine Raised inspiratory fraction of CO2 with inspiratory peak (signature capnogram with rebreathing) resulted from inhalation of CO2 rich gas during the positive pressure ventilation After a scientific literature search, we did not find such unusual reporting and its detection with the help of capnogram. “Signature capnogram” would be one of the methods to timely detect the disconnection of CO2 absorbent from anesthesia machine thereby preventing and managing intraoperative adverse respiratory events leading to catastrophe. We are reporting another valuable utility of capnography by displaying signatures capnogram in our case where soda lime canister was disconnected from the Datex ohmeda S/5 Avance machine. Further studies are required in this field.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  3 in total

1.  Failure to detect CO2-absorbent exhaustion: seeing and believing.

Authors:  D Pond; R A Jaffe; J G Brock-Utne
Journal:  Anesthesiology       Date:  2000-04       Impact factor: 7.892

Review 2.  Capnometry and anaesthesia.

Authors:  K Bhavani-Shankar; H Moseley; A Y Kumar; Y Delph
Journal:  Can J Anaesth       Date:  1992-07       Impact factor: 5.063

3.  Computer-assisted capnogram analysis.

Authors:  H R van Genderingen; N Gravenstein; J J van der Aa; J S Gravenstein
Journal:  J Clin Monit       Date:  1987-07
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.