Hairui Ren1, Zhaoyang Chen1, Xin Zhang1, Yongmei Zhao2, Zheng Wang3, Zhenglong Wu4, Haijun Xu1.
Abstract
Concentrations of S-Adenosyl-L-Methionine (SAM) in aqueous solution and fermentation liquids were quantitatively determined by surface-enhanced Raman scattering (SERS) and verified by high-pressure liquid chromatography (HPLC). The Ag nanoparticle/silicon nanowire array substrate was fabricated and employed as an active SERS substrate to indirectly measure the SAM concentration. The linear relationship between the integrated intensity of peak centered at ~2920 cm-1 in SERS spectra and the SAM concentration was established, and the limit of detections of SAM concentrations was analyzed to be ~0.1 g/L. The concentration of SAM in real solution could be predicted by the linear relationship and verified by the HPLC detection method. The relative deviations (δ) of the predicted SAM concentration are less than 13% and the correlation coefficient is 0.9998. Rolling-Circle Filter was utilized to subtract fluorescence background and the optimal results were obtained when the radius of the analyzing circle is 650 cm-1.Entities:
Year: 2016 PMID: 27818834 PMCID: PMC5081456 DOI: 10.1155/2016/4910630
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Figure 1FE-SEM images of a top view and cross-sectional side view of the SiNWA (a), (c); top view and cross-sectional side view of the Ag/SiNWA (b), (d).
Figure 2Model spectrum (black curve: 10 g/L) and the results of applying the RCF method to the curves (red curve: parameters of the analyzing circle R = 650 cm−1 and R /R = 16 cm−1/rel. unit).
Figure 3The SERS spectra of SAM solution in various concentrations (a), the quantitative relation curve of the peak intensities centered at ~2920 cm−1 of SAM (b). Each sample was measured for 5 times and the average was calculated.
Figure 4SERS spectra of SAM solution in different concentration (a). Plot for the comparison of the measured value with the calculated value (b). Each sample was measured for 5 times and the average was calculated. The SAM solution of 0 g/L was used to prove that the peak centered at ~2920 cm−1 was produced by SAM.
The calculated and measured results of 8 samples with the relative deviations.
| Sample (SAM) | S2920 | Calculated values | HPLC measured values |
|
|---|---|---|---|---|
| 1 | 147.30 | 0.14 | 0.13 | 12.60 |
| 2 | 185.76 | 0.30 | 0.31 | 0.29 |
| 3 | 206.76 | 0.39 | 0.44 | 11.13 |
| 4 | 611.42 | 2.08 | 2.08 | 0.13 |
| 5 | 1625.38 | 6.30 | 6.36 | 0.87 |
| 6 | 2221.81 | 8.78 | 8.91 | 1.36 |
| 7 | 2640.04 | 10.53 | 10.49 | 0.31 |
| 8 | 3095.71 | 12.42 | 12.79 | 2.89 |