A multicenter study to assess the reproducibility of antiphospholipid antibody results produced by an automated system.

K M Devreese1, A Poncet2, E Lindhoff-Last3, J Musial4, P de Moerloose5, P Fontana5.   

Abstract

Essentials Inter-lab variation studies for antiphospholipid antibodies (aPL) with the same assay are lacking. We carried out an assessment of repeatability and reproducibility of an automated aPL assay. High intra-center repeatability for anticardiolipin and aβ2 GPI makes duplicate testing unnecessary. Inter-lab reproducibility was high except for aβ2GPI IgG.
SUMMARY: Background Inter-assay variability is a well-known problem in antiphospholipid antibody testing, because of the lack of standardization. Inter-laboratory reproducibility for the same assay is similarly important. Objectives Testing repeatability and reproducibility of HemosIL® AcuStar for anticardiolipin (aCL) and antiβ2-glycoprotein I antibodies (aβ2GPI) IgG and IgM. Patients/Methods In this observational study, out of 420 samples from the thrombophilia centers of Ghent and Geneva, 100 samples were randomly selected and successively analyzed in three centers: Ghent (C1, in duplicate for repeatability evaluation), Geneva (C2) and Frankfurt (C3). Results Results from 99 samples were available, including 25 from patients with antiphospholipid syndrome (APS) and 74 from non-APS patients. The intra-center repeatability expressed as intra-class correlation coefficient (ICC) was higher than 0.99 for each parameter. Differences between two measurements rarely exceeded 1 U mL-1 for values below 100 U mL-1 , except for aβ2GPI IgG, where differences varied from -4 to 4 U mL-1 . The inter-center ICCs were higher than 0.99, except for aCL IgM (ICC = 0.961). These ICCs remained high even when considering values below 100 U mL-1 (0.943, 0.964 and 0.977 for aCL IgG, aCL gM and aβ2GPI IgM, respectively), except for aβ2GPI IgG (ICC = 0.652). Qualitative comparison showed less than 5% discordant classification between centers, with somewhat more discordant results for aβ2GPI IgG. Conclusions In terms of discriminating properties, the HemosIL® AcuStar has excellent intra-center repeatability and a good inter-center reproducibility for aCL IgG, aCL IgM and aβ2GPI IgM. Some concern may arise for aβ2GPI IgG.
© 2016 International Society on Thrombosis and Haemostasis.

Entities:  

Keywords:  anticardiolipin; antiphospholipid antibodies; reproducibility; β2glycoprotein I

Mesh:

Substances:

Year:  2016        PMID: 27813343     DOI: 10.1111/jth.13560

Source DB:  PubMed          Journal:  J Thromb Haemost        ISSN: 1538-7836            Impact factor:   5.824


  6 in total

Review 1.  Laboratory Diagnosis of Antiphospholipid Syndrome: Insights and Hindrances.

Authors:  Arne Vandevelde; Katrien M J Devreese
Journal:  J Clin Med       Date:  2022-04-13       Impact factor: 4.964

Review 2.  Update on Antiphospholipid Syndrome: Ten Topics in 2017.

Authors:  Ilaria Cavazzana; Laura Andreoli; Maarteen Limper; Franco Franceschini; Angela Tincani
Journal:  Curr Rheumatol Rep       Date:  2018-03-15       Impact factor: 4.592

3.  Influence of anticardiolipin and anti-β2 glycoprotein I antibody cutoff values on antiphospholipid syndrome classification.

Authors:  Laura Vanoverschelde; Hilde Kelchtermans; Jacek Musial; Bas de Laat; Katrien M J Devreese
Journal:  Res Pract Thromb Haemost       Date:  2019-05-26

4.  Role of antiphospholipid antibodies in the diagnosis of antiphospholipid syndrome.

Authors:  Katrien M J Devreese; Stéphane Zuily; Pier Luigi Meroni
Journal:  J Transl Autoimmun       Date:  2021-11-06

5.  Coagulation/Complement Activation and Cerebral Hypoperfusion in Relapsing-Remitting Multiple Sclerosis.

Authors:  Tatiana Koudriavtseva; Annunziata Stefanile; Marco Fiorelli; Caterina Lapucci; Svetlana Lorenzano; Silvana Zannino; Laura Conti; Giovanna D'Agosto; Fulvia Pimpinelli; Enea Gino Di Domenico; Chiara Mandoj; Diana Giannarelli; Sara Donzelli; Giovanni Blandino; Marco Salvetti; Matilde Inglese
Journal:  Front Immunol       Date:  2020-10-27       Impact factor: 7.561

6.  The impact of repeated freeze-thaw cycles on antiphospholipid antibody titer.

Authors:  Karel Maelegheer; Katrien M J Devreese
Journal:  Res Pract Thromb Haemost       Date:  2018-01-24
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.