| Literature DB >> 27803838 |
Jing Li1, Byung Kun Kim1, Kang-Kyun Wang1, Ji-Eun Im1, Han Nim Choi1, Dong-Hwan Kim1, Seong In Cho2, Won-Yong Lee1, Yong-Rok Kim1.
Abstract
This study demonstrates the application feasibility of electrochemical impedance spectroscopy (EIS) in measuring estrogen (17β-estradiol) in gas phase. The present biosensor gives a linear response (R2 = 0.999) for 17β-estradiol vapor concentration from 3.7 ng/L to 3.7 × 10-4 ng/L with a limit of detection (3.7 × 10-4 ng/L). The results show that the fabricated biosensor demonstrates better detection limit of 17β-estradiol in gas phase than the previous report with GC-MS method. This estrogen biosensor has many potential applications for on-site detection of a variety of endocrine disrupting compounds (EDCs) in the gas phase.Entities:
Year: 2016 PMID: 27803838 PMCID: PMC5075617 DOI: 10.1155/2016/9081375
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Scheme 1(a) Fabrication of the estrogen biosensor based on the estrogen receptor-α modified electrode and (b) the binding of estrogen hormone to the biosensor.
Figure 1(a) Nyquist plots for the faradaic impedance measurements in the presence of 5.0 mM [Fe(CN)6]3−/4− at the estrogen biosensor before (A) and after the treatment of 3.7 × 10−4 ng/L (B), 3.7 × 10−3 ng/L (C), 3.7 × 10−2 ng/L (D), 3.7 × 10−1 ng/L (E), and 3.7 ng/L (F) of 17β-estradiol in gas phase. (b) Plot of ΔR et versus −log[17β-estradiol]. R et(0): electron transfer resistance at the estrogen biosensor before the treatment of 17β-estradiol; R et(i): electron transfer resistance at the estrogen biosensor after the treatment of a certain concentration of 17β-estradiol. ΔR et = R et(i) − R et(0). (c) Four-component equivalent circuit, R et: resistance of electron transfer, R : resistance of solution, W: Warburg impedance, and CPE: constant phase element.
Values of the equivalent circuit elements for 17β-estradiol detection.
| 17 |
|
|
|
|
|---|---|---|---|---|
| 0 | 80 | 1941 | 15263 | 2479 |
| 3.7 × 10−4 | 80 | 2019 | 16982 | 3034 |
| 3.7 × 10−3 | 80 | 1917 | 19890 | 3219 |
| 3.7 × 10−2 | 80 | 1979 | 21530 | 3310 |
| 3.7 × 10−1 | 80 | 1197 | 25603 | 3895 |
| 3.7 | 80 | 2041 | 26067 | 3068 |
Figure 2Plot of [17β-estradiol]/ΔR et versus −log[17β-estradiol] in 3.7 × 10−4 ng/L to 3.7 ng/L 17β-estradiol. Inset: plot of [17β-estradiol]/ΔR et versus [17β-estradiol] in 3.7 × 10−4 ng/L to 3.7 ng/L 17β-estradiol. Values represent the mean ± standard deviation from three separate experiments. The experiments were repeated by 3 times.
Figure 3Selectivity test of the estrogen biosensor. Values represent the mean ± standard deviation from three separate experiments.