Shuyu Wang1, Shifeng Yu2, Michael S Siedler3, Peter M Ihnat4, Dana I Filoti4, Ming Lu5, Lei Zuo2.
Abstract
We developed an ultrasensitive micro-DSC (differential scanning calorimeter) for liquid protein sample characterization. This design integrated vanadium oxide thermistors and flexible polymer substrates with microfluidics chambers to achieve a high sensitivity (6 V/W), low thermal conductivity (0.7 mW/K), high power resolutions (40 nW), and well-defined liquid volume (1 μl) calorimeter sensor in a compact and cost-effective way. We further demonstrated the performance of the sensor with lysozyme unfolding. The measured transition temperature and enthalpy change were in accordance with the previous literature data. This micro-DSC could potentially raise the prospect of high-throughput biochemical measurement by parallel operation with miniaturized sample consumption.Entities:
Mesh:
Year: 2016 PMID: 27802741 DOI: 10.1063/1.4965443
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523