Effect of monocortical and bicortical screw numbers on the properties of a locking plate-intramedullary rod configuration. An in vitro study on a canine femoral fracture gap model.

Elinor J Field1, Kevin Parsons, Julie A Etches, Katie Hamilton, Neil J Burton.   

Abstract

OBJECTIVES: To evaluate the effect of varying the number and configuration of locking bicortical and monocortical screws on a plate-rod construct using a mid-diaphyseal femoral ostectomy model.
METHODS: Thirty Greyhound femurs were assigned to six groups (A-F). An intramedullary pin was placed in each bone following which a 3.5 mm locking plate was applied with six differing locking screw configurations. Groups A to C had one bicortical screw in the most proximal and distal plate holes and one to three monocortical locking screws in the proximal and distal fragments. Groups D to F had no bicortical screws placed and two to four monocortical locking screws in proximal and distal fragments. Each construct was axially loaded at 4 Hz from a preload of 10 Newtons (N) to 72 N, increasing to 144 N and 216 N, each of 6000 cycles with a further 45,000 cycles at 216 N to simulate a three to six week postoperative convalescence period. Constructs were then loaded to failure.
RESULTS: No construct suffered screw loosening or a significant change in construct stiffness during cyclic loading. There was no significant difference in load to failure of any construct (p = 0.34), however, less variation was seen with monocortical constructs. All constructs failed at greater than 2.5 times physiological load, and failure was by bending of the intramedullary pin and plate rather than screw loosening or pull-out. CLINICAL SIGNIFICANCE: Axially loaded locking monocortical plate-rod constructs applied to the canine femur may confer no difference biomechanically to those employing locking bicortical screws.

Entities:  

Keywords:  Plate-rod; bicortical locking screws; comminuted fracture; monocortical locking screws

Mesh:

Year:  2016        PMID: 27779272     DOI: 10.3415/VCOT-16-03-0040

Source DB:  PubMed          Journal:  Vet Comp Orthop Traumatol        ISSN: 0932-0814            Impact factor:   1.358


  2 in total

1.  Outcome following surgical stabilization of distal diaphyseal and supracondylar femoral fractures in dogs.

Authors:  Francisco Silveira; Isobel C Monotti; Anna M Cronin; Nick J Macdonald; Scott Rutherford; Kornelia Tiffinger; Ian Faux; Javier Rincon-Alvarez; Elvin Kulendra; Francesca Tavola; Bruno Santos; Neil J Burton
Journal:  Can Vet J       Date:  2020-10       Impact factor: 1.008

2.  Comparison between Areas of Bone Visualization Using Radiolucent Hybrid Fixator Frames and Graphically Simulated Metallic Frames: An Ex Vivo Study.

Authors:  Andrea Bonardi; Gian Luca Rovesti; Filippo Maria Martini; Francesco Dondi; Davide Benedini; Fabio Barbieri
Journal:  Vet Sci       Date:  2022-03-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.