Literature DB >> 27776886

Bending Resistance and Cyclic Fatigue Life of Reciproc, Unicone, and WaveOne Reciprocating Instruments.

Emmanuel João Nogueira Leal Silva1, Laise Silva Villarino2, Victor Talarico Leal Vieira2, Thais Accorsi-Mendonça2, Henrique Dos Santos Antunes2, Gustavo De-Deus2, Hélio Pereira Lopes3.   

Abstract

INTRODUCTION: This study evaluated the bending resistance and cyclic fatigue life of a new single-file reciprocating instrument (Unicone; Medin, Nové Město na Moravě, Czech Republic). Reciproc (VDW, Munich, Germany) and WaveOne (Dentsply Maillefer) instruments were used as references for comparison.
METHODS: Flexibility was determined by 45° bending tests using a universal testing machine. The cyclic fatigue test was performed using a custom-made device. For this test, an artificial canal with a 60° angle and a 5-mm radius of curvature was used. Scanning electron microscopic analysis was performed to determine the mode of fracture and possible deformations at the helical shaft. Statistical analysis for the bending resistance test was performed using parametric methods (ie, 1-way analysis of variance). Post hoc pair-wise comparisons were performed using the Tukey test for multiple comparisons (P < .05). Weibull analysis was used to calculate the mean life, beta, and eta parameters.
RESULTS: Reciproc presented significantly lower bending resistance than the other tested systems (P < .05), whereas no differences were observed between WaveOne and Unicone (P > .05). When mean life was compared among the brands, Reciproc lasted longer than WaveOne with a probability of 99.9%, longer than Unicone in the "RECIPROC ALL" mode with a probability of 99.9%, and longer than Unicone in the "WAVEONE ALL" mode with a probability of 99.9% (all statistically significant). Moreover, WaveOne lasted longer than Unicone in the "RECIPROC ALL" mode with a probability of 98.5% and longer than Unicone in the "WAVEONE ALL" mode with a probability of 99.8% (all statistically significant). Finally, Unicone in the "RECIPROC ALL" mode lasted longer than Unicone in the "WAVEONE ALL" mode with a probability of 95.3% (statistically significant).
CONCLUSIONS: The new reciprocating instrument Unicone showed lower cyclic fatigue resistance compared with Reciproc R25 and WaveOne Primary files.
Copyright © 2016 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bending resistance; cyclic fatigue; reciprocating movement

Mesh:

Substances:

Year:  2016        PMID: 27776886     DOI: 10.1016/j.joen.2016.08.026

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  7 in total

1.  Does the type of endodontic access influence in the cyclic fatigue resistance of reciprocating instruments?

Authors:  Emmanuel João Nogueira Leal Silva; Rafael Santos Attademo; Marcia Cristina Dutra da Silva; Karem Paula Pinto; Henrique Dos Santos Antunes; Victor Talarico Leal Vieira
Journal:  Clin Oral Investig       Date:  2020-11-21       Impact factor: 3.573

2.  Shaping ability of rotary or reciprocating systems for oval root canal preparation: a micro-computed tomography study.

Authors:  Camila Galletti Espir; Camila Almeida Nascimento-Mendes; Juliane Maria Guerreiro-Tanomaru; Bruno Cavalini Cavenago; Marco Antonio Hungaro Duarte; Mário Tanomaru-Filho
Journal:  Clin Oral Investig       Date:  2018-03-10       Impact factor: 3.573

3.  Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments.

Authors:  Emmanuel João Nogueira Leal Silva; Victor Talarico Leal Vieira; Fernanda Hecksher; Marja Ribeiro Silva Dos Santos Oliveira; Henrique Dos Santos Antunes; Edson Jorge Lima Moreira
Journal:  Clin Oral Investig       Date:  2018-02-01       Impact factor: 3.573

4.  Comparison of the Shaping Ability of WaveOne Reciprocating Files with or without Glide Path in Simulated Curved S-shaped Root Canals.

Authors:  Ayca Yilmaz; Enver Sedat Kucukay; Murat Istektepe; Selen Nihal Sisli; Handan Ersev; Isil Karagoz-Kucukay
Journal:  J Int Soc Prev Community Dent       Date:  2017-06-20

Review 5.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

6.  Cyclic Fatigue and Torsional Resistance of Four Martensite-Based Nickel Titanium Reciprocating Instruments.

Authors:  Emmanuel Silva; Carolina Oliveira de Lima; Victor Vieira; Henrique Antunes; Edson Jorge Lima Moreira; Marco Versiani
Journal:  Eur Endod J       Date:  2020-12

7.  Residual smear layer after root canal instrumentation by using Niti, M-Wire and CM-Wire instruments: A scanning electron microscopy analysis.

Authors:  Ricardo Machado; Daniel Comparin; Eduardo Donato Eing Engelke Back; Lucas da Fonseca Roberti Garcia; Luiz Rômulo Alberton
Journal:  Eur J Dent       Date:  2018 Jul-Sep
  7 in total