Alex Virgilio1, Daniela Schiavo2, Letícia M Costa3, Joaquim A Nóbrega1, Bradley T Jones4, George L Donati5.
Abstract
This work describes a procedure using the recently proposed standard dilution analysis (SDA) calibration method for the determination of As, Cr and Ni in concentrated HNO3 and HCl by inductively coupled plasma tandem mass spectrometry (ICP-MS/MS). Because of the low contaminant levels, and consequently low limits of detection required for these reagents (commonly used in trace element analysis and the semiconductor industry), samples were minimally diluted. The analysis of concentrated acids can result in matrix/transport effects, which may compromise accuracy in ICP-MS determinations. High-chlorine content samples are also challenging for As and Cr determination due to the formation of polyatomic species such as 40Ar35Cl+ and 35Cl16OH+, which overlap the only As isotope, 75As+, and the main Cr isotope, 52Cr+, respectively. The combination of SDA and ICP-MS/MS was evaluated to overcome matrix, transport and spectral interferences in order to increase accuracy, precision and sample throughput. The performance of SDA was compared with the traditional methods of external standard calibration (EC), internal standardization (IS), and standard additions (SA). The limits of detection for SDA were calculated as 6, 10, and 30ngL-1 for As, Cr, and Ni, respectively. Recoveries for spike experiments using the new method were in the 90-114% range for all analytes. The procedure described here provides similar or even better analytical performance in comparison with EC, IS and SA.Entities:
Keywords: Accuracy; Concentrated acids; High sample throughput; High-purity reagents; Tandem mass spectrometry; Two-standard calibration
Year: 2016 PMID: 27769489 DOI: 10.1016/j.talanta.2016.09.047
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057