Discrimination and quantification of two isomeric antineoplastic drugs by rapid and non-invasive analytical control using a handheld Raman spectrometer.

L M M Lê1, A Tfayli2, J Zhou2, P Prognon3, A Baillet-Guffroy2, E Caudron3.   

Abstract

Raman spectroscopy is a rapid, non-destructive and non-invasive method that is a promising tool for real-time analytical control of drug concentrations. This study evaluated a handheld Raman device to discriminate and quantify two isomeric drugs used to treat cancer. Doxorubicin (DOXO) and epirubicin (EPIR) samples were analyzed at therapeutic concentrations from 0.1 to 2mg/mL (n=90) and 0.08-2mg/mL (n=90) by non-invasive measurements using a portable Raman spectrometer. The discrimination of these two molecules was demonstrated for all concentrations (n=180) by qualitative analysis using partial least square discriminant analysis (PLS-DA) with 100% classification accuracy, sensitivity and specificity and 0% error rate. For each molecule, quantitative analyses were performed using PLS regression. The validity of the model was evaluated using root mean square error of cross validation (RMSECV) and prediction (RMSEP) that furnished 0.05 and 0.02mg/mL for DOXO and 0.17 and 0.16mg/mL for EPIR after pretreatment optimization. Based on the accuracy profile, the linearity range was from 1.256 to 2.000mg/mL for DOXO (R2=0.9988) and from 0.553 to 2.000mg/Ml for EPIR (R2=0.9240) and repeatability (CV% max of 1.8% for DOXO and 3.2% for EPIR) and intermediate precision (CV% max of 2.8% for DOXO and 4.5% for EPIR) were both acceptable. Despite the narrow validated concentration range for quantitative analysis, this study shows the potential of a handheld Raman spectrometer coupled to chemometric approaches for real-time quantification of cytotoxic drugs, as well for discriminating between two drugs with similar UV absorption profiles. Finally, the use of a handheld spectrometer with the possibility of a direct measurement of substances in containers is a potentially valuable tool for combining patient safety with security of healthcare workers.
Copyright © 2016 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Anthracycline; Discrimination; Handheld device; Quantification; Raman spectroscopy

Mesh:

Substances:

Year:  2016        PMID: 27769413     DOI: 10.1016/j.talanta.2016.07.025

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  3 in total

1.  Field detection devices for screening the quality of medicines: a systematic review.

Authors:  Serena Vickers; Matthew Bernier; Stephen Zambrzycki; Facundo M Fernandez; Paul N Newton; Céline Caillet
Journal:  BMJ Glob Health       Date:  2018-08-29

2.  Discriminative and Quantitative Analysis of Antineoplastic Taxane Drugs Using a Handheld Raman Spectrometer.

Authors:  Laetitia Lê; Marion Berge; Ali Tfayli; Patrice Prognon; Eric Caudron
Journal:  Biomed Res Int       Date:  2018-07-03       Impact factor: 3.411

3.  Chemometric Analysis of UV-Visible Spectral Fingerprints for the Discrimination and Quantification of Clinical Anthracycline Drug Preparation Used in Oncology.

Authors:  Aimen El Orche; Casimir Adade Adade; Hafid Mefetah; Amine Cheikh; Khalid Karrouchi; Miloud El Karbane; Mustapha Bouatia
Journal:  Biomed Res Int       Date:  2021-05-06       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.