Comparison of invasive and brachial cuff-based noninvasive measurements for the assessment of blood pressure amplification.

Atsushi Nakagomi1, Sho Okada1, Toshihiro Shoji1, Yoshio Kobayashi1.   

Abstract

Our aim was to assess the discrepancy in the blood pressure amplification (BPA) value defined as the aortic-to-brachial increase in systolic BP (SBP) between invasive and noninvasive brachial cuff-based methods. In 45 patients undergoing cardiac catheterization, BP in the brachial artery and ascending aorta were measured with an invasive catheter and a brachial cuff-based oscillometric device. To calculate aortic SBP, brachial waveforms were calibrated by the brachial systolic and diastolic BP (DBP) (C1 calibration) or by the brachial mean BP and DBP (C2 calibration). C1 calibration underestimated aortic SBP (-17.7 mm Hg (95% confidence interval: -21.9 to -13.5)), whereas C2 calibration generated an approximately accurate aortic SBP (1.8 mm Hg (-2.4 to 5.9)). Regarding brachial SBP, noninvasively measured values were markedly underestimated (22.2 mm Hg (-26.4 to -18.0)), resulting in a slightly low BPA value in C1 calibration (11.9±6.3 mm Hg) and a paradoxical negative BPA value in C2 calibration (-7.6±6.7 mm Hg). Multiple linear regression analysis showed that the cuff-catheter difference of BPA was positively correlated with the cuff-catheter difference of brachial SBP in both calibrations (C1 calibration: β=0.51; C2 calibration: β=0.50; both P<0.01). Although noninvasively measured BPA was associated with invasively measured BPA only in C1 calibration (r=0.33, P=0.03), when using invasively measured brachial SBP instead of a cuff-based measurement, the BPA was well associated with invasively measured BPA in both calibrations (C1 calibration: r=0.57; C2 calibration: r=0.52; both P<0.001). In conclusion, there was a trade-off in accuracy between brachial cuff-based noninvasive aortic SBP and BPA because of the inherent inaccuracies in the cuff-based method. This finding should be fully considered in establishing standardized reference values for aortic BP.

Entities:  

Mesh:

Year:  2016        PMID: 27761001     DOI: 10.1038/hr.2016.132

Source DB:  PubMed          Journal:  Hypertens Res        ISSN: 0916-9636            Impact factor:   3.872


  5 in total

1.  Systolic blood pressure amplification and waveform calibration.

Authors:  Siegfried Wassertheurer; Bernhard Hametner; James Sharman; Thomas Weber
Journal:  Hypertens Res       Date:  2017-01-19       Impact factor: 3.872

2.  Impact of Methodological and Calibration Approach on the Association of Central and Peripheral Systolic Blood Pressure with Cardiac Structure and Function in Children, Adolescents and Adults.

Authors:  Alejandro Díaz; Daniel Bia; Yanina Zócalo
Journal:  High Blood Press Cardiovasc Prev       Date:  2019-10-30

3.  A systematic review of invasive, high-fidelity pressure studies documenting the amplification of blood pressure from the aorta to the brachial and radial arteries.

Authors:  Denis Chemla; Sandrine Millasseau
Journal:  J Clin Monit Comput       Date:  2020-10-09       Impact factor: 1.977

4.  Association Between Central-Peripheral Blood Pressure Amplification and Structural and Functional Cardiac Properties in Children, Adolescents, and Adults: Impact of the Amplification Parameter, Recording System and Calibration Scheme.

Authors:  Alejandro Díaz; Daniel Bia
Journal:  High Blood Press Cardiovasc Prev       Date:  2021-02-23

5.  Twenty-Four-Hour Central (Aortic) Systolic Blood Pressure: Reference Values and Dipping Patterns in Untreated Individuals.

Authors:  Thomas Weber; Athanase D Protogerou; Mohsen Agharazii; Antonis Argyris; Sola Aoun Bahous; Jose R Banegas; Ronald K Binder; Jacques Blacher; Andréa Araujo Brandao; Juan J Cruz; Kathrin Danninger; Cristina Giannatasio; Auxiliadora Graciani; Bernhard Hametner; Piotr Jankowski; Yan Li; Alessandro Maloberti; Christopher C Mayer; Barry J McDonnell; Carmel M McEniery; Marco Antonio Mota Gomes; Annelise Machado Gomes; Maria Lorenza Muiesan; Janos Nemcsik; Anna Paini; Enrique Rodilla; Aletta E Schutte; Petros P Sfikakis; Dimitrios Terentes-Printzios; Alexandre Vallée; Charalambos Vlachopoulos; Lisa Ware; Ian Wilkinson; Robert Zweiker; James E Sharman; Siegfried Wassertheurer
Journal:  Hypertension       Date:  2021-11-15       Impact factor: 10.190

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.