Risk factors for complications in the implantation of epicardial pacemakers in neonates and infants.

A Sami Chaouki1, David S Spar2, Philip R Khoury2, Jeffrey B Anderson2, Timothy K Knilans2, David L S Morales2, Richard J Czosek2.   

Abstract

BACKGROUND: Complications related to epicardial pacemakers in infants have been reported, though limited data are available on their incidence and associated risk factors.
OBJECTIVE: The hypothesis of the study is that younger, smaller patients and larger devices would be associated with complications in neonates and infants.
METHODS: This is a retrospective study of all patients at a single center receiving an epicardial pacemaker at ≤12 months of age (1996-2015). Patient and device characteristics were obtained. Characteristics of patients with and without complications were compared.
RESULTS: There were 86 patients with a median age of 73 days (interquartile range 13-166 days), of whom 12 (14%) had a complication. Eight (9%) needed surgical intervention, of whom 5 (6%) required explantation. Younger age (9 days vs 89 days; P = .01) and lower weight (2.91 kg vs 4.44 kg; P = .004) at implantation were associated with complications. Device characteristics were not statistically different. Patients ≤3 kg in weight and/or <5 days of age had an odds ratio of 18.1 (3.6-91.2; P < .001) for developing a complication with a negative predictive value (NPV) of 97%. Regardless of weight, patients aged >21 days were found to be at lower risk with an NPV of 96%; and regardless of age, patients weighing >4 kg had an NPV of 98%.
CONCLUSION: Young age and low weight at the time of implantation are risk factors for complications, while device characteristics appear to play a minor role. Reserving pacemaker implantation for patients >3 kg in weight and 5 days of age may predict patients at low risk of developing complications.
Copyright © 2016 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Complications; Epicardial pacemakers; Infants; Pediatrics; Risk factors

Mesh:

Year:  2016        PMID: 27756705     DOI: 10.1016/j.hrthm.2016.10.017

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  Implantation of a leadless pacemaker in a pediatric patient with congenital heart disease.

Authors:  Anthony C McCanta; Gira S Morchi; Froilan Tuozo; Farhouch Berdjis; Joanne P Starr; Anjan S Batra
Journal:  HeartRhythm Case Rep       Date:  2018-08-01

2.  Video-assisted thoracoscopic pacemaker lead placement in children with atrioventricular block.

Authors:  Sergey Termosesov; Ekaterina Kulbachinskaya; Ekaterina Polyakova; Dmitriy Khaspekov; Ivan Grishin; Vera Bereznitskaya; Maria Shkolnikova
Journal:  Ann Pediatr Cardiol       Date:  2020-11-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.