L G Månsson1, S Kheddache, D Schlossman, J Börjesson, E Håkansson, S Mattsson, U Tylén.
Abstract
A digital system for chest radiography based on a large image intensifier was compared with a conventional film-screen system. The diagnostic performance was evaluated with special reference to the digital monitor images with a modified version of receiver operating characteristic (ROC) analysis--free response ROC (FROC) analysis--on a chest equivalent phantom. Measurements of spatial resolution and energy imparted were also performed. The detectability of low-contrast objects as well as spatial resolution was better for the full-size film-screen radiographs than for both the digital monitor images and the 100 mm photofluorograms. The image-intensifier system has a potential for considerable dose savings in relation to the conventional technique provided that fluoroscopy is excluded in the positioning of the patients.Entities:
Mesh:
Year: 1989 PMID: 2775592
Source DB: PubMed Journal: Acta Radiol ISSN: 0284-1851 Impact factor: 1.990