Influence of echocardiographic and radiographic characteristics on atrial sensing amplitude in patients with Linox Smart S DX defibrillation leads.

Marcin Michalak, Andrzej Cacko, Agnieszka Kapłon-Cieślicka, Monika Budnik, Przemysław Stolarz, Grzegorz Opolski, Marcin Grabowski1.   

Abstract

BACKGROUND: Single-lead for implantable cardioverter-defibrillator (ICD) with floating atrial sensing dipole is a new diagnostic tool with the potential advantage in terms of arrhythmia discrimination. We sought to determine whether right heart size and dipole position influence atrial sensing.
METHODS: Atrial sensing (AS) amplitude was measured during implantation (PP, periprocedural), predischarge (IHFU, in-hospital follow-up) and 3-6 months after the procedure (AFU, ambulatory follow-up). Results were related to atrial dipole position in the right atrium (RA) on the basis of chest X-ray examination as well as right heart dimensions at echocardiography.
RESULTS: Twenty-four patients were included into final analysis. In 14 (58.3%) patients, sensing dipole was located in regions 1 and 2 of the RA. AS amplitude was greater in regions 1 and 2 when com¬pared to other locations (3.15 vs. 1.2 mV, p = 0.045, 7.53 vs. 3.8 mV, p < 0.001 and 5.63 vs. 2.44 mV, p = 0.017 for PP measurements, IHFU and AFU, respectively). There was a significant negative correlation between AS-PP and short RA dimension (RADs) (r = -0.56, p = 0.02), AS-IHFU and RA area (RAA) (r = -0.45, p < 0.05), AS-AFU and long RA dimension (RADl) (r = -0.46; p = 0.02), AS-AFU and RADs (r = -0,48, p = 0.02), and AS-AFU and RAA (and r = -0.52, p < 0.01). There was no relationship between AS and other right heart dimensions.
CONCLUSIONS: Larger RA size and low sensing dipole location were associated with lower AS amplitude in single-lead dual chamber ICD.

Entities:  

Keywords:  DX; atrial sensing; defibrillation leads; implantable cardioverter-defibrillator

Mesh:

Year:  2016        PMID: 27747856     DOI: 10.5603/CJ.a2016.0095

Source DB:  PubMed          Journal:  Cardiol J        ISSN: 1898-018X            Impact factor:   2.737


  1 in total

1.  The efficacy of the LinoxSmart DX ICD lead from a single center experience.

Authors:  Ibrahim Marai; Anat Milman; Rafael Diamante; Osnat Gurevitz; David Barlev; Igor Lipchenka; Eyal Nof; Michael Glikson; Roy Beinart
Journal:  Indian Pacing Electrophysiol J       Date:  2019-12-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.