| Literature DB >> 27738692 |
Matthew Y H Tang1, Ho Cheung Shum2.
Abstract
We present a wash-free high-sensitivity immunoassay of C-reactive proteins with droplet microfluidics. Microbeads are encapsulated within droplets for the immunoassay, and the droplets are scanned by a fluorescence detection platform to quantify the amount of proteins captured on the microbeads. The limit of detection determined by our platform is 0.01 μg mL-1, which is ten times more sensitive than conventional high-sensitivity C-reactive protein assays. With the decrease in diffusion distance within droplets, the immunoassay requires only half of the time required for similar conventional approaches. This approach for carrying out immunoassays can potentially be applied to other biomarkers beyond C-reactive proteins.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27738692 DOI: 10.1039/c6lc01121g
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799