Literature DB >> 27731958

Case study on human α1-antitrypsin: Recombinant protein titers obtained by commercial ELISA kits are inaccurate.

Henning Gram Hansen1, Helene Faustrup Kildegaard1, Gyun Min Lee1,2, Stefan Kol1.   

Abstract

Accurate titer determination of recombinant proteins is crucial for evaluating protein production cell lines and processes. Even though enzyme-linked immunosorbent assay (ELISA) is the most widely used assay for determining protein titer, little is known about the accuracy of commercially available ELISA kits. We observed that estimations of recombinant human ø1-antitrypsin (rø1AT) titer by Coomassie-stained SDS-PAGE gels did not correspond to previously obtained titers obtained by a commercially available ELISA kit. This prompted us to develop two independent quantification assays based on biolayer interferometry and reversed-phase high-performance liquid chromatography. We compared the rø1AT titer obtained by these assays with three different off-the-shelf ELISA kits and found that the ELISA kits led to inconsistent results. The data presented here show that recombinant protein titers determined by ELISA kits cannot be trusted per se. Consequently, any ELISA kit to be used for determining recombinant protein titer must be validated by a different, preferably orthogonal method.
© 2016 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

Entities:  

Keywords:  Alpha-1 antitrypsin; Biolayer interferometry; Chinese hamster ovary (CHO) cells; ELISA; RP-HPLC

Mesh:

Substances:

Year:  2016        PMID: 27731958     DOI: 10.1002/biot.201600409

Source DB:  PubMed          Journal:  Biotechnol J        ISSN: 1860-6768            Impact factor:   4.677


  1 in total

1.  A reversed phase HPLC method for the quantification of HIV gp145 glycoprotein levels from cell culture supernatants.

Authors:  José A González-Feliciano; Coral M Capó-Vélez; Pearl Akamine; Manuel Delgado-Vélez; Ruth Almodóvar; Javier Rivera; Ignacio Pino; Gloriner Morell; Daniel Eichinger; José H Rivera; José A Lasalde-Dominicci; Abel Baerga-Ortiz
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2021-01-27       Impact factor: 3.205