Arterialization of central venous pressure waveform.

Monish S Raut1, Arun Maheshwari1, Manish Sharma2, Sandeep Joshi1, Arun Kumar1, Akshay Gupta3, Himanshu Goyal3.   

Abstract

Entities:  

Mesh:

Year:  2016        PMID: 27716718      PMCID: PMC5070347          DOI: 10.4103/0971-9784.191556

Source DB:  PubMed          Journal:  Ann Card Anaesth        ISSN: 0971-9784


× No keyword cloud information.
The Editor, A 60-year-old female known diabetic and hypertensive presented with fever since 5 days. She was diagnosed to have urinary tract infection with left ventricular systolic dysfunction. Noradrenaline infusion was started to maintain hemodynamics. Right subclavian vein was punctured, and triple lumen central venous catheter (CVC) was inserted for infusion of vasopressor drugs. Central venous pressure (CVP) monitoring was attached to CVC. Monitor showed unusual CVP waveform - resembling arterial upstroke and downstroke [Figure 1]. However, waveform tracing was not showing peak pressure more than 20 mmHg. Noninvasive blood pressure in the right arm was 104/66 mmHg. Blood gas analysis of blood sample derived from CVC confirmed venous origin of blood. Why was there arterial waveform in venous catheter? Chest X-ray of the same patient revealed course of the right subclavian venous catheter into right internal jugular vein (IJV) [Figure 2]. Ultrasound of right side of the neck showed pulsating internal carotid artery (confirmed by ultrasound Doppler) just beside IJV with CVC [Figure 3]. Arterial pulsation of carotid artery might be transmitting to IJV, which was reflected in waveform [Video 1].
Figure 1

Monitor display showing central venous pressure waveform resembling arterial waveform

Figure 2

Chest X-ray showing the course of central venous catheter from subclavian to internal jugular vein

Figure 3

Ultrasound image showing internal jugular vein with central venous catheter adjacent to internal carotid artery. Left panel - internal carotid artery with Doppler waveform. Right panel - internal jugular vein with Doppler waveform

Monitor display showing central venous pressure waveform resembling arterial waveform Chest X-ray showing the course of central venous catheter from subclavian to internal jugular vein Ultrasound image showing internal jugular vein with central venous catheter adjacent to internal carotid artery. Left panel - internal carotid artery with Doppler waveform. Right panel - internal jugular vein with Doppler waveform CVP waveform normally consists of three positive waves (a, c, and v waves) and two negative waves (x and y descents). Abnormalities in these waves can be produced by varieties of clinical cardiac conditions.[12] Venous catheter can give arterial waveform as it travels through right ventricle to pulmonary artery, for example, Swan-Ganz catheter. However, in the present case, venous catheter gave arterial waveform even in the absence of it's entry into cardiac structures or artery. CVC showing airway pressure waveform due to catheter migration has been reported.[3] Arterial waveform showed by a CVC generally indicates inadvertent arterial insertion.[4] Accidental position of CVC in the left internal thoracic vein or the pericardiophrenic vein mimicking as arterial waveform has also been reported.[5] The author concluded the “arterial” waveform by CVC was due to transmission of pulse waves of adjacent cardiac contraction and filling.[5] Waveform analysis, chest X-ray, and blood gas interpretation can diagnose misplacement of CVC. Pulsatile vascular structure such as artery in the immediate vicinity of vein with catheter can give arterial waveform due to transmissible pulsation. The present case demonstrated arterial waveform in CVC does not always imply arterial placement of catheter.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  3 in total

1.  Airway pressure waveform transduced from central venous catheter indicating catheter migration.

Authors:  Ashish Bangaari
Journal:  Anesth Analg       Date:  2010-04-01       Impact factor: 5.108

2.  Arterial and central venous pressure monitoring.

Authors:  Atilio Barbeito; Jonathan B Mark
Journal:  Anesthesiol Clin       Date:  2006-12

3.  Inadvertent arterial insertion of a central venous catheter: delayed recognition with abrupt changes in pressure waveform during surgery -A case report-.

Authors:  Yong Sun Choi; Ji Young Park; Young Lan Kwak; Jong Wha Lee
Journal:  Korean J Anesthesiol       Date:  2011-01-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.