Literature DB >> 2771361

Calibration of Hirschberg test photographs under clinical conditions.

P A DeRespinis1, E Naidu, S E Brodie.   

Abstract

The feasibility of determining the angle of ocular deviation from clinical photographs was assessed. Corneal reflex photographs were obtained from 30 strabismic patients and 5 orthophoric control subjects, using a centered electronic flash, with a millimeter ruler placed across the forehead for calibration. Measurements of corneal reflex displacement were obtained from the photographs and compared with standard prism-cover results. The value of 20.89 prism diopters (PD) per millimeter of corneal reflex displacement was obtained for the factor of conversion from reflex displacement to strabismic angle. This value of the Hirschberg coefficient is valid only for measurements of reflex displacement along a true frontal plane, as provided in a photograph. Detailed algorithms for the calculation of strabismic angle from photographic data are provided in an appendix.

Entities:  

Mesh:

Year:  1989        PMID: 2771361     DOI: 10.1016/s0161-6420(89)32775-8

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  4 in total

1.  Mathematical analysis and experiment on the corneal reflex test in spectacle wearers.

Authors:  Y Hatsukawa; H Koike; S Sugimoto; K Kasai; P Fells
Journal:  Br J Ophthalmol       Date:  1994-09       Impact factor: 4.638

2.  Variability of Ocular Deviation in Strabismus.

Authors:  John R Economides; Daniel L Adams; Jonathan C Horton
Journal:  JAMA Ophthalmol       Date:  2016-01       Impact factor: 7.389

Review 3.  Tests for detecting strabismus in children aged 1 to 6 years in the community.

Authors:  Sarah Hull; Vijay Tailor; Sara Balduzzi; Jugnoo Rahi; Christine Schmucker; Gianni Virgili; Annegret Dahlmann-Noor
Journal:  Cochrane Database Syst Rev       Date:  2017-11-06

4.  The reliability of the angle of deviation measurement from the Photo-Hirschberg tests and Krimsky tests.

Authors:  S Tengtrisorn; A Tungsattayathitthan; S Na Phatthalung; P Singha; N Rattanalert; S Bhurachokviwat; S Chouyjan
Journal:  PLoS One       Date:  2021-12-01       Impact factor: 3.240