Literature DB >> 2770636

Dose parameters of 125I and 192Ir seed sources.

K A Weaver1, V Smith, D Huang, C Barnett, M C Schell, C Ling.   

Abstract

As mandated by an NCI brachytherapy contract, we measured dosimetric parameters for 192Ir seeds and two models of 125I seeds. Measurements were with LiF powder in a water-equivalent phantom. Data were corrected for background, sample mass, and finite detector volume. Selected parameters were also investigated through Monte Carlo calculations. Results are presented in terms of a dose parametrization that is described in detail, and are compared to published data. Our results agreed well with published data for relative quantities such as radial and angular dose dependence. Our measured value for the 192Ir dose factor was 4.55 cGy(H2O) cm2 mCi-1 h-1, also in good agreement with commonly used values. However, the measured dose factors for 125I seed models 6702 and 6711 were 1.18 and 1.06 cGy(H2O) cm2 mCi-1 h-1, values well below those in general use.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2770636     DOI: 10.1118/1.596322

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Microwave plaque thermoradiotherapy for choroidal melanoma.

Authors:  P T Finger
Journal:  Br J Ophthalmol       Date:  1992-06       Impact factor: 4.638

2.  Dosimetric parameters of three new solid core I-125 brachytherapy sources.

Authors:  Timothy D Solberg; John J DeMarco; Geoffrey Hugo; Robert E Wallace
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

3.  Measured transverse-axis dosimetric parameters of the model STM1251 125I interstitial source.

Authors:  Z Li; J F Williamson
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

4.  Optical fibre based real-time measurements during an LDR prostate brachytherapy implant simulation: using a 3D printed anthropomorphic phantom.

Authors:  P Woulfe; F J Sullivan; L Byrne; A J Doyle; W Kam; M Martyn; S O'Keeffe
Journal:  Sci Rep       Date:  2021-05-27       Impact factor: 4.379

  4 in total