Abstract
To decrease health-risks to humans, non-toxic compounds were evaluated for use as internal standards for calibrating data obtained by gas chromatography/multiphoton ionization mass spectrometry (GC-MPI-MS) using an ultraviolet femtosecond laser as the ionization source. The retention time in the mass chromatogram was calibrated using a retention index, in which a series of n-alkanes was employed as internal standards for evaluating the retention times for polychlorinated dibenzo-p-dioxins (PCDDs), polychlorinated dibenzofurans (PCDFs), and polychlorinated biphenyls (PCBs). To compensate for changes in signal intensity in MPI-MS, the dependence of signal intensity on the laser pulse energy was investigated for the dioxin-like compounds, in addition to five non-toxic aromatic hydrocarbons, that were used as internal standards. Based on their similar behavior,the non-toxic PCDD/PCDF, its 13C-isotope, and pentachlorobenzene behave similarly, we conclude that they can be used for calibrating the signal intensities in MPI-MS.Entities:
Keywords: Dioxin; Internal standard; Mass spectrometry; Multiphoton ionization
Mesh:
Substances:
Year: 2016 PMID: 27702616 DOI: 10.1016/j.chroma.2016.09.071
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759